Err
or Codes
ClosureRFG Radiofrequency Generator
7-7
123
Temperature and
Power are too high.
Impedance is too low.
Cycle system power and allow the POST to complete. If the
error persists, insert a known, working catheter and cycle
system power. If the system does not respond and/or the
same error code is displayed, contact Covidien customer
service.
124
Impedance and Power
are too high.
125
Temperature is too
low. Impedance and
Power are too high.
126
Temperature,
Impedance, and Power
are too high.
201
Temperature is too
high.
202
Temperature is too
low.
203
Impedance is too high.
204
Impedance is too low.
205
Power is too high.
206
Power is too low.
314
Task took too long to
complete.
Cycle system power and allow the POST to complete. If the
system does not respond and/or the same error code is
displayed, contact Covidien customer service.
350
Temperature exceeds a
safety limit.
Cycle system power and allow the POST to complete. If the
error persists, insert a known, working catheter. Cycle
system power again. If the system does not respond and/or
the same error code is displayed, contact Covidien customer
service.
352
The catheter power
exceeded the
ClosureRFG
radiofrequency
generator Power
Alarm Limit.
354
The catheter voltage
exceeded the
ClosureRFG generator
Voltage Alarm Limit.
358
The catheter was
disconnected while in
Measure or Treatment
Mode.
Error
Code
Description
Corrective Action