Chapter 6
ClosureRFG Radiofrequency Generator
6-1
Messages, Tones, and Responses
Condition and System Responses
Name
Action
User Action
High Temperature
Advisory
Display: “Advisory: Temperature high”
Sound: Informational Tone.
None required
Low Temperature Advisory
Display: “Advisory: Temperature low”
Sound: Informational Tone.
None required
High Impedance Advisory
Display: “Advisory: Impedance high”
Sound: Informational Tone.
None required
Low Impedance Advisory
Display: “Advisory: Impedance low”
Sound: Informational Tone.
None required
Low Power Advisory
Display: “Advisory: Power low” Sound:
Informational Tone.
None required
High Power Advisory
Display: “Advisory: Power high” Sound:
Informational Tone.
None required
High Impedance
Functional Advisory
Display: “Advisory: Impedance high.
Check Connection.” Sound:
Informational Tone.
None required
Low Impedance Functional
Advisory
Display: “Advisory: Impedance low.
Replace catheter.”Sound: Informational
Tone.
None required
High Power And High
Temperature Advisory
Clear any advisory message. Display:
“Advisory: Temperature high, power
high” Sound: Informational Tone.
None required
Low Power And High
Temperature Advisory
Clear any advisory message. Display:
“Advisory: Temperature high, power
low” Sound: Informational Tone.
None required
High Impedance And High
Temperature Advisory
Clear any advisory message. Display:
“Advisory: Temperature high” Sound:
Informational Tone.
None required