7-2
ClosureRFG Radiofrequency Generator
8
Resource conflict.
Cycle system power and allow the POST testing to complete.
If the system does not respond and/or the same error code is
displayed, contact Covidien customer service.
21
Memory failure.
22
Memory failure.
23
UIC watchdog failure
24
Operator-stored
settings invalid.
Default settings
restored.
Go to the Settings screen on the device and press the
Defaults button. Cycle system power. If the system does not
respond and/or the same error code is displayed, contact
Covidien customer service.
25
The inserted USB
device does not
contain valid
information or is
missing files.
Cycle system power without the USB device installed. Allow
the system to boot and then insert the USB device. If the
error persists, use a known, valid USB device. If the system
does not respond and/or the same error code is displayed,
contact Covidien customer service.
26
The inserted USB
device cannot upgrade
this ClosureRFG
generator because its
serial number does not
match.
Cycle system power without the USB device installed. Allow
the system to boot and then insert the USB device. If the
error persists, attempt to use a known, valid USB device.
Verify that the serial number on the rear of the device
matches the serial number given on the About tab of the
Settings menu. If the serial numbers do not match and/or the
same error code is displayed, contact Covidien customer
service.
28
The inserted USB
device is not correctly
formatted or is an
invalid type.
Cycle system power without the USB device installed. Allow
the system to boot and then insert the USB device. If the
error persists, attempt to use a known, valid USB device.
Reformat the USB device by right clicking the device and
selecting “format” on your PC. The device must be
formatted to FAT32. If the system does not respond and/or
the same error code is displayed, contact Covidien customer
service.
29
The ClosureRFG
radiofrequency
generator is unable to
write data to the USB
device.
30
The inserted USB
device is full and no
more data can be
stored on it.
Remove the USB device, clear the data on the device and
reinstall. Cycle system power and allow POST to complete.If
the issue persists, try a different USB device. If the system
does not respond and/or the same error code is displayed,
contact Covidien customer service.
31
The real-time clock
stopped and has been
restarted.
Go to the Settings screen on the device and press the
Defaults button. Cycle system power. After POST, re-enter
the Settings screen and adjust the date and time as
explained in
Setting System Time
on page 2-6. If the system
does not respond and/or the same error code is displayed,
contact Covidien customer service.
Error
Code
Description
Corrective Action