Chapter 10
ClosureRFG Radiofrequency Generator
10-1
Service & Maintenance
Maintenance
Important
There is no scheduled maintenance for the ClosureRFG radiofrequency generator. There are no
calibrations. The ClosureRFG generator self-calibrates during POST.
The ClosureRFG generator needs no routine service other than cleaning and service activity
required by the operator’s institution. Qualified personnel should periodically inspect the
ClosureRFG generator, and if service is necessary, contact Covidien.
Warning
The ClosureRFG generator contains no operator-serviceable parts. The case should not be
removed. The ClosureRFG generator must be returned to the factory for service.
No modification of this equipment is allowed.
General Safety Guidelines
Review
General Safety Guidelines
on page 1-6.
In addition:
• The AC power switch on the front panel serves as the means of disconnection from
supply mains. This switch also isolates the ClosureRFG generator electrically from the
supply mains on both poles simultaneously.
• Using a power cord that does not meet the specifications in this Service and
Maintenance section may present a potential for an electric shock or affect the
performance of the ClosureRFG generator.
Environmental Conditions
-4°F
-20°C
158°F
70°C
90%
10%
-4°F 104°F
-20°C 40°C
to
60%
10%
104°F 158°F
40°C 70°C
to
Store the ClosureRFG radiofrequency generator at temperatures between
-20° C and 40° C and non-condensing relative humidity between 10% and 90% or
temperatures between 40° C and 70° C and non-condensing relative humidity between
10% and 60%.
50°F
10°C
104°F
40°C
90%
10%
Operate the ClosureRFG generator at temperatures between 10 º C and 40 º C
and in non-condensing relative humidity between 10% and 90%.