Chapter 7
ClosureRFG Radiofrequency Generator
7-1
Error Codes
If an error occurs, the ClosureRFG radiofrequency generator displays an error code. An
error is one of two types: unrecoverable or recoverable.
•
Unrecoverable
errors stop the ClosureRFG generator from functioning and require
the operator to acknowledge the condition. Once the operator presses the green
check-mark, the ClosureRFG generator (itself or with assistance) cycles power to
continue. If the problem persists, contact Covidien Customer Service or a local
distributor.
•
Recoverable
errors require the operator to respond to an error message or make a
clinical decision. After acknowledgement, the ClosureRFG generator returns to the last
viewed screen.
Error codes
Error
Code
Description
Corrective Action
1
Unexpected state.
Cycle system power and allow the POST testing to complete.
If the system does not respond and/or the same error code is
displayed, contact Covidien customer service.
2
Unexpected event.
3
A parameter value was
incorrect.
4
A queue overflowed.
5
Unit not responding.
A communication cable within the unit has failed or is not
installed correctly. Cycle power on the system and allow the
POST testing to complete. If the system does not respond
and/or the same error code is displayed, contact Covidien
customer service.
6
Software version
incorrect.
Cycle system power and insert a USB device with the latest
SW. Upgrade the system using the About tab. Cycle system
power. If the system does not respond and/or the same error
code is displayed, contact Covidien customer service
7
Memory corruption
detected.
Cycle system power and allow the POST testing to complete.
If the system does not respond and/or the same error code is
displayed, contact Covidien customer service.