ClosureRFG Radiofrequency Generator
RF Power Delivery Performance
220-240 VAC, 5 A
• Cable: H05VVF3G1.0 VDE, maximum length 10 ft. (3 m)
• Plug: minimum 6 A - 240 VAC
• Unit receptacle: IEC female, minimum 6 A - 240 VAC
Conditions for Transporting and Storing the ClosureRFG
Generator
•
-4°F
-20°C
158°F
70°C
Temperature between -20° C and 70° C
•
90%
10%
-4°F 104°F
-20°C 40°C
to
60%
10%
104°F 158°F
40°C 70°C
to
Relative humidity (non-condensing) between 10% and 90% for
temperatures between -4° F and 104° F (-20° C and 40° C), and between 10% and
60% for temperatures between 104° F and 158° F (40° C and 70° C).
•
500 hPa
1060 hpa
Ambient atmospheric pressure between 500 hPa and 1060 hPa
Conditions for Operating the ClosureRFG Generator
•
50°F
10°C
104°F
40°C
Temperature between 10° C and 40° C
•
90%
10%
Relative humidity (non-condensing) between 10% and 90%
•
700 hPa
1060 hPa
Ambient atmospheric pressure between 700 hPa and 1060 hPa (equivalent to
elevation between 3000 m and -300 m)
RF Power Delivery Performance
The fundamental frequency of the RF signal is set at 460.8 kHz.
Measurement Tolerances
Displayed Value
Range
Accuracy
Time
0:00 -99:59 minutes:seconds
±1s
Temperature
10° - 130° C
±5° C
Power*
0.1 - 40 W
±20% (@ 25-1000 Ω)
Impedance
10 - 1999 Ω
±15%
or ±5 Ω, whichever is greater
* RF power is measured on the isolated patient circuit and is the average power. The RF power
capacity is also limited by the Maximum Power (W) setting of the connected RF catheter.
Note:
Refer to the RF catheter’s instructions for use for more information.