Chapter 5
ClosureRFG Radiofrequency Generator
5-1
Treatment Times
Retain Total Treatment Timer—Connecting a New
Catheter
When Used
• An RF treatment has been completed, either CLF or RFS.
• The catheter was unplugged and reconnected, or a new one of the same type
connected.
• Power was not cycled on the ClosureRFG radiofrequency generator.
When a catheter is disconnected and the same or identical one reconnected, it will be used
on either the same patient/patient site, or a new patient/patient site. The screen shows
this question:
Would you like to retain the total treatment timer value?
?
1
Would you like to retain the total treatment timer value?
How
•
Same patient or site:
To continue adding to the previous timer value, press the
Green
Check Mark
button. The ClosureRFG generator will behave as if the catheter had
never been unplugged.
•
Different patient or site:
To reset the timer to zero, press the
Red X
button. This
closes the current record (stores the accumulated treatment time and cycles), and
opens a new data record for the next RF treatment.