ClosureRFG Radiofrequency Generator
6-5
Audible Signals
Audible Signals
The ClosureRFG radiofrequency generator produces a audible signal for out-of-limit
conditions based on the power, temperature, and impedance measurements.
During multiple message or tone conditions, the message or tone with the highest priority
takes precedence, as shown in the following two tables.
Treatment Summary
Export Failed
Display: “Treatment summary data
export failed.”
Use another USB
device
USB Device Full
Display: “Device full. Please replace with
another USB device.” “No records
exported.”
Use another USB
device
USB Not Installed
Display: “USB device is not installed.”
“No records exported.”
Insert USB device
ClosureFast Treatment
Ramp Up
Display: “Device should reach target
temperature in XX seconds.” Output RF
ramp up tone.
None required
Confirm Discard
Display: “Would you like to retain the
total treatment timer value?” Wait for
User Acknowledgement.
User
acknowledgement
Error Log Corruption
Display: “Error Log Corrupted. All error
entries have been erased. Press OK to
restart system.” Wait for User
Acknowledgement.
User
acknowledgement.
Generator restarts.
Non Recoverable Error
Display: “Error #ww. Please press
to
restart.” in the message area. Display:
the error code only as part of this
message. Display: Standby screen in the
Data Area. Wait for User
Acknowledgement. Restart generator.
User
acknowledgement.
Generator restarts.
Recoverable Error
Display: “Error #ZZ. Press
to restart.”
in the message area. Display: the error
code only as part of this message.
Display: Standby screen in the Data Area.
Sound: Informational Tone. Wait for
User Acknowledgement.
User
acknowledgement.
Unsupported Catheter
Display: “The connected device is
unsupported. Please connect another
device.”
Replace catheter
Name
Action
User Action