
4-2
ClosureRFG Radiofrequency Generator
Ready-to-Treat Screen
ClosureRFS Stylet Gauges
Data Display with Gauges
Gauges indicate the measured parameter and whether the stylet’s parameters
(temperature, impedance, and RF power) are within acceptable ranges. If a parameter is
outside the advisory limits, its gauge brightens within the red area or lights only the blue
area, a message is displayed, and a tone is sounded.
ClosureRFS gauges during treatment
ClosureRFS
225
Ω
6.0
w
s
0:10
85
°c
1
2
3
Illumination of the
temperature
gauge segments:
•
Blue:
Cool, waiting for, or at the beginning of a heating cycle; below advisory limit
during treatment
•
Blue-Green:
Within normal range (after initial heating)
•
Red:
Temperature above advisory or functional limit
ཱ
Location of the bright
impedance
gauge segments:
•
Red-yellow (low end):
Below advisory or functional limit
•
Blue-Green:
Within normal range to advisory
•
Red:
Above advisory or functional limit
ི
Illumination of the
Power
gauge segments:
•
Blue and Blue-Green:
Within normal range to advisory
•
Red:
Above advisory or functional limit