6-3
Condition and System Responses
High Power Low
Temperature And High
Impedance Advisory
Clear any advisory message. Display:
“Advisory: Temperature low. Power
high. Impedance high” Sound:
Informational Tone.
None required
High Power High
Temperature And High
Impedance Advisory
Clear any advisory message. Display:
“Advisory: Temperature high. Power
high. Impedance high” Sound:
Informational Tone.
None required
High Temperature
Functional Limit
Clear any advisory message. Display:
“Treatment Halted: Temperature high.
Please press
to continue.” Sound:
Audio Medium Alarm. Wait for User
Acknowledgement.
User
acknowledgement
Low Temperature
Functional Limit
Clear any previous Advisory condition.
Display: “Treatment Halted:
Temperature low. Please press
to
continue.” Sound: Informational Tone.
Wait for User Acknowledgement.
User
acknowledgement
High Impedance
Functional Limit
Display: “Treatment Halted: Impedance
high. Please press
to continue.”
Sound: Informational Tone. Wait for
User Acknowledgement.
User
acknowledgement
Low Impedance Functional
Limit
Display: “Treatment Halted: Impedance
low. Please press
to continue.” Sound:
Informational Tone. Wait for User
Acknowledgement.
User
acknowledgement
High Power Functional
Limit
Display: “Treatment Halted: Power high.
Please press
to continue.” Sound:
Audio Medium Alarm. Wait for User
Acknowledgement.
User
acknowledgement
Low Power Functional
Limit
Display: “Treatment Halted: Power low.
Please press
to continue.” Sound:
Informational Tone. Wait for User
Acknowledgement.
User
acknowledgement
Catheter Disconnected
Display: “Treatment Halted: Device
disconnected. Please press
to
continue.” message. Sound:
Informational Tone. Wait for User
Acknowledgement.
User
acknowledgement
Advisory Catheter Broken
Display: “The connected device is
broken. Please connect another device.”
Sound: Informational Tone. Wait for
User Acknowledgement.
User
acknowledgement.
Replace catheter.
Name
Action
User Action