Chapter 9
ClosureRFG Radiofrequency Generator
9-1
Technical Specifications
Equipment Type
For ClosureRFG radiofrequency generator repairs or returns, call the Covidien Customer
Service Department. The ClosureRFG generator may be returned to the factory.
The Class I ClosureRFG radiofrequency generator is designed to work with Type CF,
Defibrillator-Proof RF catheters. The ClosureRFG generator is designed to withstand an
application of an external defibrillator while the RF catheter is in use.
Electrical: ClosureRFG Generator
• Operates from an AC power source ranging from 100 V to 240 VAC, 50 Hz to 60 Hz,
5A.
• Includes a DIN 42801–compatible equipotential grounding lug that meets the ground
bond requirements of UL 60601–1.
• USB: 2.0 devices only
Should the supplied power cord need to be replaced to match another plug configuration,
the replacement plug/cable/receptacle configuration must meet or exceed the following
specifications:
100-120 VAC, 5 A
• Cable: SJT16/3, IEC color code, maximum length 10 ft. (3 m)
• Plug: minimum 10 A - 125 VAC
• Unit receptacle: IEC female, minimum 10 A - 125 VAC