5-2
ClosureRFG Radiofrequency Generator
Treatment Time Reset—Using the Same Catheter
Treatment Time Reset—Using the Same Catheter
When Used
• An RF treatment has been done, either CLF or RFS.
• The catheter will be reused at a different patient site.
• The ClosureRFG radiofrequency generator was not power cycled.
RF treatment sometimes occurs at more than one site in a patient. If the operator wants
to use the same catheter, and also wants a record of the treatment time at each site, the
record for one site must be closed, and a new record for the next site opened.
How
To reset the RF treatment Timer to zero (and open a new record)
1. Press the
Reset Timer Value
button
at the bottom of the screen. The screen
prompts for confirmation.
Would you like to retain the total treatment timer value?
?
1
Would you like to retain the total treatment timer value?
2. Press the
Green Check Mark
button to reset the RF treatment timer, and open a new
record.
or
Press the
Red X
button to cancel the reset, and continue with the current record.
The timer will also be reset to zero and a new record opened with the following steps:
a. Unplug the catheter and then reconnect, or connect another catheter of the same
type.
b. The screen prompts, “Would you like to retain the total treatment timer value?”
c. Press the
Red X
button to not retain the Treatment Timer, and start a new
treatment record. This will perform the reset.
Note:
Cycling power on ClosureRFG generator will also reset the timer and close
the current record.