10-4
ClosureRFG Radiofrequency Generator
Disposal
Touch Screen:
The recommended agent for cleaning the surface is IPA in water, 70/30.
Alcohol wipes and prep pads are acceptable.
Allow 15 minutes after cleaning for the excess cleaning agent to evaporate before turning
on the ClosureRFG generator.
ClosureRFG generator surfaces
(other than the touch screen): The recommended
agents for cleaning, to prevent surface degrading or discoloration, are:
• Five percent solution of household bleach (approximately 0.25% Sodium
Hypochlorite). Disinfecting wipes or pads that contain only Sodium Hypochlorite are
acceptable.
• IPA in water, 70/30.
All surfaces, including the bottom side of the ClosureRFG generator and any surface on
which it is placed, must be cleaned of any residues.
The ClosureRFG generator cannot be sterilized and must not enter a sterile surgical field.
Disposal
Follow local governing ordinances and recycling plans regarding disposal or recycling of
components.
Important
The ClosureRFG radiofrequency generator includes a lithium battery. Do not incinerate or dispose
of lithium batteries in general trash collection. Check state and local regulations dealing with the
disposal of these materials.