Err
or Codes
ClosureRFG Radiofrequency Generator
7-3
32
The inserted USB
device is not
formatted.
Cycle system power without the USB device installed. Allow
the system to boot and then insert the USB device. If the
error persists, attempt to use a known, valid USB device.
Reformat the USB device by right clicking the device and
selecting “format” on your PC. The device must be
formatted to FAT32. If the system does not respond and/or
the same error code is displayed, contact Covidien customer
service.
34
A UIC Software Task
did not check in with
the Task Monitor.
Cycle system power and allow the system POST to complete.
If the system does not respond and/or the same error code is
displayed, contact Covidien customer service.
35
The ClosureRFG
generator has
experienced more than
five resets in one
minute.
Unplug the system’s power cord from the wall. Re-insert the
plug into the wall receptacle and allow the system POST to
complete. If the system does not respond and/or the same
error code is displayed, contact Covidien customer service.
36
A Software Stack
became corrupt.
Cycle system power and allow the POST to complete. If the
error persists, insert a USB device with the latest system SW.
Upgrade the system from the About tab. Cycle system
power. If the system does not respond and/or the same error
code is displayed, contact Covidien customer service.
37
The temperature of the
ClosureRFG generator
is above the specified
operating
temperature.
Cycle system power allow the POST to complete. If the error
persists, insert a known, working catheter and cycle system
power again. Should the internal temperature of the device
remain too high and the error persists, disconnect the power
cord from the power source. Allow 30 minutes for the device
to cool to ambient temperature. Re-attach the system to the
power source and allow the unit to complete the POST. If the
system does not respond and/or the same error code is
displayed, contact Covidien customer service.
38
The temperature of the
ClosureRFG generator
is below the specified
operating
temperature.
Cycle system power and allow the POST testing to complete.
If the error persists, insert a known, working catheter. Cycle
the system power again. Should the internal temperature of
the device remains too low and the error persists, allow the
system to sit in an on state for a minimum of 30 minutes. If
the system does not respond and/or the same error code is
displayed, contact Covidien customer service.
39
Real Time Clock
Failure.
Cycle system power and allow the POST to complete. If the
system does not respond and/or the same error code is
displayed, contact Covidien customer service.
Error
Code
Description
Corrective Action