7-6
ClosureRFG Radiofrequency Generator
106
Impedance is too high.
Cycle system power and allow the POST to complete. If the
error persists, insert a known, working catheter and cycle
system power. If the system does not respond and/or the
same error code is displayed, contact Covidien customer
service.
107
Temperature is too low
and Impedance is too
high.
108
Temperature and
Impedance are too
high.
109
Power is too low.
110
Temperature and
Power are too low.
111
Temperature is too
high and Power is too
low.
112
Impedance and Power
are too low.
113
Temperature,
Impedance, and Power
are too low.
114
Temperature is too
high. Impedance and
Power are too low.
115
Impedance is too high
and Power is too low.
116
Temperature and
Power are too low.
Impedance is too high.
117
Temperature and
Impedance are too
low. Power is too high.
119
Power is too high.
120
Temperature is too
low. Power is too high.
121
Impedance is too low
and Power is too high.
122
Temperature and
Impedance are too
low. Power is too high.
Error
Code
Description
Corrective Action