10 ORIGINAL (ORG) MODE
10.2.3 Memory Switch Table
No.
Setting
Function
00
0
DET stage Judgement timing 1)
1
01
0
1
02
0
External stop: b contact pulse
1
External stop: a contact pulse
03
0
After the tester makes judgement, the judgement signals are held.
These signals are cleared by the START, STOP or CHG HOLD signals.
1
The judgement signal is output as pulse.
04
0
Intelligent I Pneumatic Circuit: AV check will not be performed
1
Intelligent I Pneumatic Circuit: AV check will be performed
05
0
Accepts START signal during the END stage.
1
Does not accept START signal during the END stage.
06
0
Exhaust Interference Prevention is activated.
Tester goes to DL3 and exhausts when START signal is turned off.
1
Tester goes to DL3 and exhausts right after judgement is made.
07
0
Offset error is not detected in DL1 stage.
1
Offset error is detected in DL1 stage.
08
0
Busy signal (control I/O port pin#36) 2)
1
09
0
1
10
0
1
11
0
±300 Pa for large leak limit in the DL2 stage.
*1
1
±1000 Pa for large leak limit in the DL2 stage.
12
0
CHG stage display 3)
1
13
0
1
14
0
The judgement signal is transmitted at END stage.
*2
1
The judgement signal is transmitted at DL3 stage.
15
0
RESERVED
1
RESERVED
16
0
Pilot pressure is monitored.
Set to 1 for other than the
pressure switch built-in unit.
*3
1
Pilot pressure is not monitored.
17
0
Test pressure is not monitored with upper and lower limits.
1
Test pressure is monitored with upper and lower limits.
18
0
Baud rate: 19,200 bps
1
Baud rate: 9600 bps
19
0
Serial communication output format selection 4)
1
20
0
1
21
0
RESERVED
1
RESERVED
22
0
Test pressures of odd-numbered channels are monitored with OUT1 and even-numbered
channels are monitored with OUT2.
1
Checks the test pressure only with OUT1of pressure switch.
23
00 to
34
Pressurization valve malfunction 5)
24
0
Change the maximum values of the CHG timer and the STB timer to 6000 seconds.
*4
1
Change the maximum values of the CHG timer and the STB timer to 999.8 seconds.
*1
For the micro-pressure specification, large leak may not be large enough for ±1000 Pa. Select ±300
Pa for large leak limit in such cases. If the leak is judged to be large, “999 (-999)” and “HI NO GO (LO
NO GO)” will blink on the LEAK indicator.