11 INTERFACE
77
4) Response
Command from host computer
Response of LS-1866
Buzzer
Read command
Data transmission
OFF
Write and control command
ACK(06H) transmission
OFF
Ineffective command or
command while tester is busy
Error transmission
ON
•
Error response
#SS_00_CC_EE:GG<CR>
Data field
Symbol
Data type
Unit
Min.
Max. Remarks
Identification No.
S
2 digit integer
00
99
Channel #
C
2 digit integer
00
15
Error field
E
2 digit integer
01
80
Checksum
G
2 digit integer
00
FF
Data from “#” to “:” is summed up
and its two’s complement is taken.
•
Error field
Code
Error
Description
01
Inappropriate data
Written value was inappropriate
10
Execution not available
Tester is not in a state to execute the command.
40
Checksum error
Checksum data does not agree.
80
Ineffective command
Command is inappropriate.
•
Transmitted data from LS-1866 by RLD command (Output format is same as T format.)
#SS_00_J_±VV.VV:GG<CR>
Data field
Symbol
Data type
Unit
Min. Max. Remarks
Identification No.
Symbol
2 digit integer
00
99
Judge field
J
ASCII code
(1 digit
hexadecimal)
0
D
0: No test data
1:Lo NG
2:GOOD
4:Hi NG
9:LL NG
C:HH NG
D:ERROR
Data
V
4 digit floating
point
Leak
rate unit
±0.000
±0999.
Checksum
G
2 digit integer
00
FF
Data from “#” to “:” is summed up
and its two’s complement is taken.
•
Transmitted data from LS-1866 by other reading command
#SS_00_CC_±VVVV.VVV:GG<CR>
Data field
Symbol
Data type
Unit
Min.
Max.
Remarks
Identification No.
S
2 digit integer
00
99
Channel #
C
2 digit integer
00
15
Data
V
Fixed point
Leak
rate unit
±0000.000
±9999.000
Checksum
G
2 digit integer
00
FF
Data from “#” to “:” is
summed up and its two’s
complement is taken.
11.5.2 Output formats
1) Operation
By the setting of memory SW, various settings can be made.
2) Baud rate setting
By the setting of memory SW_#18, the baud rate can be changed.
MSW 18
1
9600
0
19200
The channel # for the transmitted data for RTO is always “00” because
RTO cannot be programmed independently for each channel.
NOTE