64
10 ORIGINAL (ORG) MODE
4) Serial communication output format
Memory switch # Serial communication output format
20
19
0
0
RESERVED
0
1
Printer format
1
0
I format
1
1
T format
5) Selection of Pressurization valve failure detection time
・
Differential pressure limit
Memory switch#23
Time
Differential pressure limit
00
OFF
OFF
11
1sec
±100Pa
12
1sec
±300Pa
13
1sec
±500Pa
14
1sec
±1000Pa
21
5sec
±100Pa
22
5sec
±300Pa
23
5sec
±500Pa
24
5sec
±1000Pa
31
10sec
±100Pa
32
10sec
±300Pa
33
10sec
±500Pa
34
10sec
±1000Pa
Pressurization valve failure detection
Verifies DPS offset during a pre-determined waiting time, upon completion of
the leak test, If DPS offset exceeds its limit, error message (E-9) is displayed
by regarding the status of AV1 (pressurization valve) as not closed.
NOTE
Memory switch #23 needs to be set at “00” when the positive test pressure is
less than 10kPa or the negative one is less than –20kPa.
NOTE