7 BASIC OPERATIONS
7.11.2 Viewing raw output of detected differential pressure
Press the RIGHT
►
key for 1 second or longer while tester is in Measurement mode, and the raw output of
detected differential pressure is displayed on the leak rate indicator.
This feature is convenient when you want to know the detected differential pressure in case the detected
value is displayed in leak rate unit mL/min, or the detected value is compensated.
7
7
.
.
1
1
2
2
T
T
i
i
m
m
e
e
r
r
S
S
e
e
t
t
t
t
i
i
n
n
g
g
s
s
7.12.1 Viewing
<Operation>
1) Press the SHIFT key to display
SFt
on the leak rate indicator.
2) Press the TIME/LIMIT key to display a stage symbol on the leak indicator.
3) Press the
◄
LEFT or RIGHT
►
key to switch the stage to be setup. A timer setting will be displayed
on both HI and LO leak limit indicators.
4) When the Master-Preset feature is off, each time you press the RIGHT
►
key the display will change in
the following sequence:
dL1
→
CHG
→
dL2
→
bAL
→
dEt
→
DL3
→
End
Press the
◄
LEFT key to move backwards.
5) When the Master-Preset feature is on, each time you press the RIGHT
►
key the display will change in
the following sequence:
dl1
→
CHG
→
dL2
→
bAL
→
dEt
→
STB
→
DL3
→
End
Press the
◄
LEFT key to move backwards.
7.12.2 Entry
A timer setting will be displayed on both HI and LO leak limit indicators.
If the keyboard is unlocked, the least significant digit of the display blinks.
<Operation>
1) Use the
▲
UP and DOWN
▼
keys to change the
setting.
Press the
▲
UP key to increment the value by one.
Press the DOWN
▼
key to decrement the value by one.
Press the SHIFT key to move the current setup digit to
left.
2) Keep depressing the
▲
UP or DOWN
▼
key increments
or decrements the value continuously.
3) After changing a setting, press the ENTER key to complete or press STOP to cancel it. Store the
value in the memory and display the next stage timer.
4) Press the STOP key to return to Measurement mode.
If you press the ENTER key in the CHG state, the STB timer will be
automatically set to the same time as CHG . If you want to set the STB timer
to a time different from the CHG timer, set it in the STB state.
NOTE