IVAC® Volumetric Pump (Models 571 & 572)
58/104
1000SM00018 Issue 2
Spare Parts Replacement Procedures
Body Assembly
(continued)
Door Sensor Assembly
Replacement Procedure
1. Unfasten the one retaining clip which secures the door sensor assembly to the inside of the front panel.
2. Disconnect the wiring harness (J11) from the Accessory Interface Board, removing cable ties as required.
3. Withdraw the door sensor assembly.
4. Reassemble in reverse order.
Refitting note:
Insert the door sensor assembly through the front panel, mount on peg and secure with clip.
Spare Parts
Item
Description
Part Number
A
FSTNR,SPD NUT LKG 0.250 SSTUD
302902
B
ASSY,DOOR SNSR BD,56X/57X(BOM)
132536
(
A
) Fastener Clip
(
B
) Door Sensor Assembly
Cable Tie
To provide a clear view of the fitting area, the
illustration below shows the door removed. The door is
not
serviceable.
If a replacement door is required, contact your local
Cardinal Health Affiliate Office or Distributor for advice.