IVAC® Volumetric Pump (Models 571 & 572)
71/104
1000SM00018 Issue 2
Spare Parts Replacement Procedures
Door Assembly
(continued)
Pressure Plate
Replacement Procedure
1. Mark the face of the pressure plate to ensure it can be refitted with the correct top side upwards.
2. Remove the four screws and washers securing the pressure plate to the door.
3. Remove the pressure plate and two coil springs seated in the door recesses.
4. Reassemble in reverse order.
Refitting notes:
1) Install the pressure plate with the correct top side upwards using the mark made on face when removed.
2) Use Loctite 243 on screws when refitting the pressure plate.
Spare Parts
Item
Description
Part Number
A
SPR,CPRSN,.360X.813X.813X.035,MW
302900
B
WSHR,FL,.187X.438X.020,BLK,SS
301594
C
SCREW,MACH 8-32X3/8 PNH PHH
301522
D
ADHESIVE LOCTITE 243
0000ME00672
(
A
) Spring (x2)
Pressure Plate
(
B
) Washer (x4)
(
C
/
D
) Screw (x4) / Loctite
Important:
Each pressure plate is mated to the cam followers during manufacture burn-in. If the pressure
plate is damaged or a replacement is required, contact your local Cardinal Health Affiliate Office or
Distributor for advice.
Correct top-side-up orientation of the pressure plate must be observed when it is refitted.
Summary of Contents for IVAC 571
Page 1: ...IVAC Variable Pressure Volumetric Pump Models 571 572 Technical Service Manual...
Page 37: ...Chapter 4 Troubleshooting In this chapter Introduction 38 Error Messages 38 General Faults 42...
Page 97: ...Appendix C Configured Options...
Page 99: ...Appendix D Disposal...
Page 101: ...Appendix E Service Contacts...
Page 103: ...Appendix F Document History...