IVAC® Volumetric Pump (Models 571 & 572)
19/104
1000SM00018 Issue 2
Configuration & Calibration
Dry Set Pressure Calibration (Test 13)
continued
12. Press the
J
unit and
RATE
switches simultaneously for a minimum of one second to set the 000 mmHg (000 cmH2O)
calibration point.
13. Reload the IV infusion set and apply test pressure until the meter reads
400
mmHg (544 cmH2O). When the pressure
reading is stable, press the
TOT VOL INF
and the
J
100 switches simultaneously for a minimum of one second to set
the 400 mmHg (544 cmH2O) calibration point.
14. Press the
RUN/HOLD
switch to return to test
13
and turn the pump OFF. Remove the IV infusion set (first bleeding off
any pressure).
15. Disconnect the battery pack from the battery connector.
16. Remove the test jumper cable and reassemble the pump.
17. Perform the 'Dry Set Pressure Verification' procedure as detailed in Chapter 3, 'Routine Maintenance'.
Calibration Procedures
(continued)
Both Steps 12 and 13 must be performed to correctly set the calibration parameters. Performing
only one step will result in the message RECONFIG REQD when the pump returns to operation.