IVAC® Volumetric Pump (Models 571 & 572)
8/104
1000SM00018 Issue 2
Introduction and Start Up
Controls and Indicators
UP/DOWN
Increases or decreases infusion rate and pressure and volume values. Three
columns for faster/slower increase or decrease:
Left = 100's, Centre = 10's, Right = units (or in micro mode Left = 10's,
Centre = units, Right = fractions).
TOT
VOL
INF
CLEAR
Clears total volume infused when both switches are held down together.
TOT
VOL
INF
TOTAL VOLUME INFUSED
Displays total volume infused value (p secondary volume infused).
Sets the Information Display to continuously monitor volume infused.
LIMIT
PRES
PRESSURE LIMIT
Displays or changes the occlusion pressure alarm level limit.
READ
PRES
READ PRESSURE
Displays the IV infusion site pressure.
Sets the Information Display to continuously monitor infusion pressure.
PRI
SEC
PRIMARY/SECONDARY
(Model 572 only). Switches between PRIMARY and SECONDARY infusion
modes. Allows displays of VTBI or rate setting of one mode while operating
in another.
CRIS
CRIS
On Model 572: only used in Self-test mode.
On Model 571: hidden switch, only used in Self-test mode.
RATE
RATE
Changes the infusion rate.
VOL
TO BE
INF
VOLUME TO BE INFUSED
(VTBI)
Sets the value of Volume To Be Infused. Displays volume remaining to be
infused.
RUN
HOLD
RUN/HOLD
Starts and stops pump infusion. Silences/cancels alarms and advances
start-up instruction messages.
ON
OFF
ON/OFF
Turns the pump on and off.
AC POWER INDICATOR
When illuminated, indicates the pump is connected to an AC power supply
and the battery is being charged.
BATTERY INDICATOR
Flashes to indicate the pump is operating on battery power.
0
With a Flow Sensor in use, an illuminated ("0") in the Information Display window indicates a drop is
detected by the flow sensor in the drip chamber.
Flashing Displays When the pump is operating on battery power, both Displays flash on/off.
Front Panel
The English switch panel controls and indicators are described below. For information on overlays in other languages,
refer to the relevant DFU. See also Appendix B, 'Spare Parts Listing' for switch panel part numbers.
Rear Panel
PRESSURE UNIT SELECTOR
SWITCH
Sets the displayed pressure units. Millimetres of mercury (mmHg) or
Centimeters of water (cmH20).
053(
4/
2%3%4
RESET SWITCH
Resets the circuit breaker.
Summary of Contents for IVAC 571
Page 1: ...IVAC Variable Pressure Volumetric Pump Models 571 572 Technical Service Manual...
Page 37: ...Chapter 4 Troubleshooting In this chapter Introduction 38 Error Messages 38 General Faults 42...
Page 97: ...Appendix C Configured Options...
Page 99: ...Appendix D Disposal...
Page 101: ...Appendix E Service Contacts...
Page 103: ...Appendix F Document History...