Configuration & Calibration
Self-test mode is used to access a number of configuration and test routines which are designed to test and verify many
of the pump functions, defaults and calibrations without requiring internal inspection. Configuration settings are also
accessed via self-test mode.
Each test is accessed by selecting a Test ID number as shown in the table below.
Self-Test Mode
Test ID
Description
001
Software Revision Display
002
Configuration Status Display
003
Flash Display Test
004
Segment Display Test
005
Sequence Display Test
006
EEPROM Test and Data Display
007
A/D Channels Test
008
Rate Calibration Display/Set
009 *
Configuration Rate Display/Set
010
Maximum Rate Display/Set
011 *
VTBI Configuration Display/Set
012
Pressure Display
013
Pressure Calibration Set
014
Maximum Occlusion Pressure Limit Display/Set
015
Default Occlusion Pressure Limit Display.Set
016
Alarm History Display/Clear
017
Switch Test
018
AC Power Run Time Display/Reset
019
Battery Run Time Display/Reset
020
Secondary Mode Completion Tone Display/Set
021
Air-In-Line Test
022
I/O Port Test
023
Language Display/Set
024
CIM ID Display/Set
025
Alarm Tone Display/Set
026*
AIL Alarm Threshold Display/Set
027
System Vacuum Display
028
Enter Operational Setup Mode
029
Burn-In Test
030
Not used
031*
Infusion Mode Configuration Display/Set
* A mode select jumper (part number 132350) must be
used to change settings, see 'Mode Select Jumper' for
further details.
Exiting Self-Test Mode
Self-test mode can be exited at any time:
Turn the pump off by pressing the
ON/OFF
switch
OR
Select test
028
then press the
RUN/HOLD
switch
to enter the start-up mode of operation
Entering Self-Test Mode
1. Press the
RUN/HOLD
and
ON/OFF
switches
simultaneously until the display turns on. Initially,
the current software revision level is displayed.
2. Press the
RUN/HOLD
switch. The Information
Display will read:
TEST ID
and the Numeric
Display will read:
002
which means the pump is
ready to start test
002
.
Note: The display may bypass the display of the
software revision level and advance directly to
TEST ID 002
, depending on how long the
RUN/
HOLD
switch is held down.
3. Use the
J
K
switches to select the Test ID
required.
Test Execution
1. After selecting the Test ID number (see above),
press the
RUN/HOLD
switch to start the test.
2. To exit a test, press the
RUN/HOLD
switch.
The three columns of
J
K
switches
allow faster/slower increments:
Left = 100's
Centre = 10's
Right = Units
Summary of Contents for IVAC 571
Page 1: ...IVAC Variable Pressure Volumetric Pump Models 571 572 Technical Service Manual...
Page 37: ...Chapter 4 Troubleshooting In this chapter Introduction 38 Error Messages 38 General Faults 42...
Page 97: ...Appendix C Configured Options...
Page 99: ...Appendix D Disposal...
Page 101: ...Appendix E Service Contacts...
Page 103: ...Appendix F Document History...