IVAC® Volumetric Pump (Models 571 & 572)
72/104
1000SM00018 Issue 2
Spare Parts Replacement Procedures
Door Assembly
(continued)
Door Latch
Replacement Procedure
1. Drive out the groove pin using a pin punch.
2. Remove the two curved spring washers and the door latch.
3. Remove the magnet cover and magnet as required.
4. Reassemble in reverse order.
Spare Parts
Item
Description
Part Number
A
PIN,GRV 0.125X1.25 S ALLOY
302053
B
COVER,MAGNET
125337
C
MAGNET,MAGNETIZED,CO 90C98A
302425
D
ADHESIVE LOCTITE 496
0000ME00753
E
ASSY LATCH,GNRC,(BOM)
132777
E
ASSY,LATCH,ENG,(BOM)
132776 (with label)
F
SPRING,CVR,WSHR U-138-0045S
301591
Refitting notes:
1) When refitting the magnet, ensure the south pole faces outwards, secure magnet in place using loctite and cover.
2) Fit the two curved spring washers into the latch recesses outwardly, convex.
3) Fit the groove pin flush with the outside surface of the door.
(
B
) Cover
(
A
) Groove pin
(
C/D
) Magnet / Loctite
(
F
) Spring Washers (x2)
(
E
) Latch
See
Inside
View
Inside View
When refitting the
door latch, apply
lub grease 33
(0000ME00452) on
a cotton swab, to
entire length of slot
inside the latch.
Summary of Contents for IVAC 571
Page 1: ...IVAC Variable Pressure Volumetric Pump Models 571 572 Technical Service Manual...
Page 37: ...Chapter 4 Troubleshooting In this chapter Introduction 38 Error Messages 38 General Faults 42...
Page 97: ...Appendix C Configured Options...
Page 99: ...Appendix D Disposal...
Page 101: ...Appendix E Service Contacts...
Page 103: ...Appendix F Document History...