IVAC® Volumetric Pump (Models 571 & 572)
73/104
1000SM00018 Issue 2
Spare Parts Replacement Procedures
Door Assembly
(continued)
Air-In-Line Transmitter Assembly
Replacement Procedure
1. Disconnect the air-in-line cable harness connector (J4) from the Display Board.
2. Remove the two screws securing the air-in-line transmitter to the door.
3. Remove the air-in-line transmitter assembly by pulling it away from the door, drawing out the cable harness.
4. Reassemble in reverse order.
Spare Parts
Item
Description
Part Number
A
P,GRV 0.125X.500 REV TPR
302269
B
ASSY,AIL,TRANS,571/572(BOM)
137269
C
SCREW,CAP,4-40X1/4,BUTH,HXSO
304172
Door Latch Pin
1. Pull out the old pin with pliers.
2. Reassemble in reverse order.
(
A
) Pin
0.25 inches ± 0.15
(6.35 ± 3.81 mm)
Front Panel
Pin
Refitting the Door Latch Pin
(
B
) Air-In-Line Transmitter
(
C
) Screw (x2)
Summary of Contents for IVAC 571
Page 1: ...IVAC Variable Pressure Volumetric Pump Models 571 572 Technical Service Manual...
Page 37: ...Chapter 4 Troubleshooting In this chapter Introduction 38 Error Messages 38 General Faults 42...
Page 97: ...Appendix C Configured Options...
Page 99: ...Appendix D Disposal...
Page 101: ...Appendix E Service Contacts...
Page 103: ...Appendix F Document History...