IVAC® Volumetric Pump (Models 571 & 572)
55/104
1000SM00018 Issue 2
Spare Parts Replacement Procedures
Chassis Assembly
(continued)
Battery Housing, Speaker, Rear Panel
Replacement Procedure
1. Disconnect the speaker harness from the Power Regulator Board (J2).
2. Remove the speaker assembly by breaking it loose from the battery housing.
3. Remove the two nuts securing the battery housing to the rear panel assembly.
4. Remove the CE panel and rear panel as required.
5. Reassemble in reverse order.
Refitting note:
Use Loctite 38050 to fix the speaker in place.
Spare Parts
A
ASSY,SPEAKER,56X/57X(BOM)
125425
B
NUT,KEP 6-32 S PL
300036
C
HSG,BATT,56X/57X
125318
D
KIT,CE PANEL,571/2
136853
E
ASSY, PANEL REAR ,RFI,571/572XB
137851
Speaker harness
(
C
) Battery Housing
(
B
) Nut (x2)
(
A
) Speaker Assembly / Loctite 38050
(
D
) CE Panel
(
E
) Rear Panel