Routine Maintenance
Dry Set Pressure Verification (Test 12)
When servicing the pump perform the following pressure calibration verification procedure.
Note: If the pump fails this test, perform the 'Dry Set Pressure Calibration' procedure. See Chapter 2. for instructions.
Procedure:
1. Connect the pump to the AC mains and set up the equipment as shown below..
Note: Set the pressure unit selector switch (on the back panel of the pump) to
mmHg
.
2. Enter self-test mode, press the
RUN/HOLD
switch and select test
12
. Press the
RUN/HOLD
switch again to display the
pressure
MMHG
.
Allow the pump to warm up for approximately 30 minutes before continuing with this test.
3. Load an IV infusion set and close the door.
4. Apply a pressure of
000
mmHg and verify that the pump displays a reading of
-20
to
+10
mmHg.
5. Apply a pressure of
400
mmHg and verify that the pumps displays a reading of
367
to
423
mmHg.
#
Reservoir
Pressure
Gauge
Air
Pressure Source
Proximal
End of Set
Do not allow the applied test pressure to exceed the intended pressure value by more than 50
mmHg. If this occurs, reload the IV infusion set and repeat test
The pressure read accuracy must be ±15 mmHg or ± 7%, whichever is greater
When applying the 000 mmHg pressure, ensure the set has just been loaded. If necessary reload
the IV infusion set.
Summary of Contents for IVAC 571
Page 1: ...IVAC Variable Pressure Volumetric Pump Models 571 572 Technical Service Manual...
Page 37: ...Chapter 4 Troubleshooting In this chapter Introduction 38 Error Messages 38 General Faults 42...
Page 97: ...Appendix C Configured Options...
Page 99: ...Appendix D Disposal...
Page 101: ...Appendix E Service Contacts...
Page 103: ...Appendix F Document History...