IVAC® Volumetric Pump (Models 571 & 572)
85/104
1000SM00018 Issue 2
Specifications
Electromagnetic Compatibility
Warning:
•
The use of any accessory, transducer, or cable with the IVAC® Volumetric Pump other than those specified may result in
increased emissions or decreased immunity of the pump.
•
The IVAC® Volumetric Pump should not be used adjacent to or stacked with other equipment, however if adjacent
or stacked use is necessary, the IVAC® Volumetric Pump should be observed to verify normal operation in the
configuration in which it will be used.
Caution:
•
The IVAC® Volumetric Pump is a CISPR 11 Group 1 Class A Medical Equipment System and intended for use by
healthcare professionals only.
•
Medical Electrical Equipment needs special precautions regarding EMC and needs to be installed, put into service and
used according to the EMC information provided in the accompanying documents.
•
Portable and Mobile RF communications can affect Medical Electrical Equipment.
•
Operating the pump near equipment which radiates high energy radio frequencies (electro surgical or cauterizing
equipment, portable radios, cellular telephones, etc.) may cause false alarm conditions. If this happens, reposition the
pump away from the source of interference or turn off the pump and manually regulate the flow.
Guidance and Manufacturer’s Declaration – Electromagnetic Emissions
The IVAC® Volumetric Pump is intended for use in the electromagnetic environment specified below.
The customer or the user of the IVAC® Volumetric Pump should assure that it is used in such an environment.
Emissions Test
Compliance
Electromagnetic Environment – Guidance
CISPR 11
RF Emissions
Group 1
The pump uses RF energy only for its internal function in the normal
product offering. Therefore, its RF emissions are very low and are not
likely to cause any interface in nearby electronic equipment.
CISPR 11
RF Emissions
Class A
The pump is suitable for use in all establishments, other than
domestic, and those directly connected to the public low-voltage
power supply network that supplies buildings used for domestic
purposes.
EN 61000-3-2
Harmonic Emissions
Class A
EN 61000-3-3
Voltage Fluctuations,
Flicker Emissions
Complies
Important: The IVAC® Volumetric Pump has been assessed to the following standards;
EN/IEC 60601-1-2:1993 (Models 571XAX, 572XAX & 572XBX)
EN/IEC60601-1-2:2001 (Models 571XBX & 572XCX).
Summary of Contents for IVAC 571
Page 1: ...IVAC Variable Pressure Volumetric Pump Models 571 572 Technical Service Manual...
Page 37: ...Chapter 4 Troubleshooting In this chapter Introduction 38 Error Messages 38 General Faults 42...
Page 97: ...Appendix C Configured Options...
Page 99: ...Appendix D Disposal...
Page 101: ...Appendix E Service Contacts...
Page 103: ...Appendix F Document History...