IVAC® Volumetric Pump (Models 571 & 572)
25/104
1000SM00018 Issue 2
Routine Maintenance
Self-Test Functions
(continued)
Test
ID
Test
Action/Description
017
Switch Test
Information Display: Initially shows
SWCH TEST
. Press each switch and verify the
corresponding code is displayed:
Switch:
Display code:
Switch:
Display code:
TOT VOL INF
PRI
RATE
LIMIT PRES
SEC
VTBI
READ PRES
TL V INF
PRI
RATE
LIMT PRS
SEC
VTBI
READ PRES
CRIS
J
100
K
100
J
10
K
10
J
Unit
K
Unit
CRIS
UP 100
DN 100
UP 10
DN 10
UP 1
DN 1
Notes:
1) The
ON/OFF
switch does not have a code.
2) The Numeric Display remains blank.
018
Read Hours on
AC
Read/clear the total number of hours that the pump has run on AC power.
Information Display:
AC HRS
Numeric Display: Number of hours on AC power.
To reset the total AC power hours to
0000
:
Press and hold the
READ PRES
switch and the
J
10 switch for at least 2 seconds.
019
Read Hours on
Battery
Read/clear the total number of hours that the pump has run on battery power.
Information Display:
BATT HRS
Numeric Display: Number of hours on battery power.
To reset the total AC battery hours to
0000
:
Press and hold the
READ PRES
switch and the
J
unit switch for at least 2 seconds.
021
Air-In-Line Test
Verify air-in-line sensor operation. To run test:
Load an air-filled or fluid-filled IV infusion set.
Information Display:
00000000
= fluid detected in IV infusion set and door is closed.
10000000
= air detected in IV infusion set or door is open.
Numeric Display:
UAIL
022
I/O Port Test
Read data on the various input ports. Press the
J
and
K
switches to select each port.
Information Display: Shows the 8 input port data bits in binary format.
Numeric Display: Indicates which port is being examined:
000
= display port, Display U7.
008
= switch port, Display U5.
016
= Logic U20.
048
= A/D end
of conversion, U22.
064
= Logic U21.
080
and
081
= parallel I/F ports.
096
,
097
,
098
and
099
= serial I/F ports.
128
= timer of Logic U14.
129
= Logic U23.
130
= Logic U24.
Key troubleshooting tests:
Port
016
bit 0 = rotation sensor 1
bit 1 = rotation sensor 2
bit 2 = set sensor
bit 3 = air-in-line sensor
Port
064
bit 7 = door sensor
027
System Vacuum
Test
Read the vacuum level being applied by the vacuum system when an IV infusion set is loaded
and the door is closed.
Information Display:
VAC MMHG
Numeric Display: Current vacuum value that is generated to hold the pressure sensing disc in
place.
To run the test, following the instructions in 'Vacuum Retention Verification' later in this
chapter.
Summary of Contents for IVAC 571
Page 1: ...IVAC Variable Pressure Volumetric Pump Models 571 572 Technical Service Manual...
Page 37: ...Chapter 4 Troubleshooting In this chapter Introduction 38 Error Messages 38 General Faults 42...
Page 97: ...Appendix C Configured Options...
Page 99: ...Appendix D Disposal...
Page 101: ...Appendix E Service Contacts...
Page 103: ...Appendix F Document History...