IVAC® Volumetric Pump (Models 571 & 572)
59/104
1000SM00018 Issue 2
Spare Parts Replacement Procedures
Body Assembly
(continued)
Air-In-Line Receiver Assembly
Replacement Procedure
1. Remove the two screws securing the air-in-line receiver assembly to the front panel.
2. Disconnect the wiring harness from the Accessory Interface Board (J1) and pull the wiring harness free of the silicone
fixing on the inside of the front panel.
3. Remove the air-in-line receiver assembly, drawing the wiring harness out through the front panel hole..
4. Reassemble in reverse order.
Spare Parts
Item
Description
Part Number
A
SCREW,CAP,4-40X1/4,BUTH,HXSO
304172
B
ASSY,AIL,RECEIVER,57X(BOM)
133996
(
A
) Screw (x2)
(
B
) Air-In-Line Receiver
Refitting note:
Secure the wiring
harness with
adhesive silicone,
RTV 162 as shown
Front Panel Back View
To provide a clear view of the fitting area, the
illustration below shows the door removed. The door is
not
serviceable.
If a replacement door is required, contact your local
Cardinal Health Affiliate Office or Distributor for advice.