IVAC® Volumetric Pump (Models 571 & 572)
16/104
1000SM00018 Issue 2
Configuration & Calibration
Test ID
Setting
Action/Data Displayed
015
Default Occlusion
Pressure
DEFAULT
Display and/or set the default pressure limit.
Pressure values are displayed in
MMHG
or
CMH20
depending on the pressure
units selected. See 'Selecting Pressure Units' for further information.
This setting determines the occlusion pressure limit at power up.
To change the setting:
Press and hold the
PRI
(upper hidden switch on Model 571) while using the
J K
switches to adjust the value.
Press the
RUN/HOLD
switch to save the new value.
020
Secondary Mode
Completion Tone
(Model 572 only)
Display and/or set the audible tone that sounds when the SECONDARY infusion
finishes and switches to PRIMARY infusion.
1 CLICK
= one-click tone
4 BEEPS
= four-beeps tone
To change the setting:
Press the
J
unit switch to select
4 BEEPS
or the
K
unit switch to select
1
CLICK
.
023
Language
Display and/or set the language to be displayed. Languages are
ENGLISH,
DUTCH, FRENCH, GERMAN, ITALIAN, SPANISH
or
SWEDISH
.
To change this setting:
Press the
J K
unit switches to scroll through list and select language.
024
CIM ID
CIM ID
Setting is not in use.
OFF
= CIM not in use,
001
to
999
= CIM ID number
Important: Ensure the setting is
OFF
.
To change setting to
OFF
:
Press and hold the
VOL TO BE INF
switch while using the
J K
switches to
select
OFF
.
Press the
RUN/HOLD
switch to save the new setting.
025
Alarm Tone
Display and/or select the alarm tone volume level.
Approximate decibel volumes are:
HI TONE = 90 db
MED TONE = 80 db
LOW TONE = 70 db
To change the setting:
Press the
J
unit switch to select alarm tone.
WARNING: Take any necessary precautions against prolonged exposure to the
HI TONE
alarm as this may result in hearing damage.
026
AIL Alarm Threshold
UL AIR
Display and/or set the air bubble length maximum value. This value is the
nominal air bubble threshold in microlitres used for air-in-line detection. Value
must be between 50 and 150 microlitres.
To change the setting:
Install the mode select jumper. See 'Mode Select Jumper'.
Press the
CRIS
(lower hidden switch on Model 571) and
VOL TO BE INF
switches together and wait for a click.
Press the
J K
switches to enter the value (in 10 µl increments).
031
Infusion Mode
INFM CFG
Display and/or set the pump infusion mode.
PRI ONLY
= PRIMARY infusion mode only (Model 571).
SEC/CRIS
= PRIMARY and SECONDARY infusions modes. (Model 572).
To change the setting:
Install the mode select jumper. See 'Mode Select Jumper'.
Press the
CRIS
(hidden switch on Model 571) and
VOL TO BE INF
switches
together and wait for a click. Press the
J
unit switch to select
PRI ONLY
or
the
K
unit switch to select
SEC/CRIS
.
Note: Model 571 must be set to
PRI ONLY.
Configuration Settings
(continued)
Summary of Contents for IVAC 571
Page 1: ...IVAC Variable Pressure Volumetric Pump Models 571 572 Technical Service Manual...
Page 37: ...Chapter 4 Troubleshooting In this chapter Introduction 38 Error Messages 38 General Faults 42...
Page 97: ...Appendix C Configured Options...
Page 99: ...Appendix D Disposal...
Page 101: ...Appendix E Service Contacts...
Page 103: ...Appendix F Document History...