IVAC® Volumetric Pump (Models 571 & 572)
15/104
1000SM00018 Issue 2
Configuration & Calibration
In order to perform certain tests and to change any pump settings, a mode select jumper must be installed.
Procedure:
1. With the pump off, remove the test connector cover from the back of the rear case.
2. Connect the mode select jumper (part number 132350) to the test connector, orientating the jumper so that CR1 is at
the top.
3. Place the pump in self-test mode then select and carry out the test(s) as required.
4. When the test is complete, switch the pump off and remove the mode select jumper from the test connector.
5. Refit the test connector cover.
Mode Select Jumper
Configuration Settings
Test ID
Setting
Action/Data Displayed
009
Rate Configuration
RATE CFG
Display and/or set the current rate range.
STANDARD
= rate range of 001 to 999 ml/h
ALL RATE
= rate range of 0.1 to 999.9 ml/h. Note: Use this setting where
fractional flow rates are required.
To change the setting:
Install the mode select jumper. See 'Mode Select Jumper'.
Press the
CRIS
(lower hidden switch on Model 571) and
VOL TO BE INF
switches together and wait for a click.
Press the
J
unit (right column) switch to select
STANDARD
or the
K
unit
switch to select
ALL RATE
.
010
Maximum Rate
MAX RATE
Display and/or set the maximum infusion rate.
999
= set to STANDARD rate configuration (ml/h). This rate cannot be changed.
0.1 and 999.9
= set to ALL RATE configuration (ml/h).
Note: Rate configuration is set in Test
009
.
To change the maximum rate setting (for ALL RATE configuration):
Press and hold the
PRI
(upper hidden switch on Model 571) while using the
J K
switches to adjust the value.
011
VTBI Configuration
VTBI CFG
Display and/or set the Volume To Be Infused (VTBI) limit. VTBI limit ranges are:
MACRO = 1 to 9999 ml, MICRO = 0.1 to 999.9 ml.
If set to ALL RATE configuration (see Test 009),
VTBI CFG
can be set to
MACRO
or
MICRO
.
If set to STANDARD rate configuration,
VTBI CFG
is set to
MACRO
and is not
adjustable.
To change the VTBI limit (for ALL RATE configuration):
Install the mode select jumper. See 'Mode Select Jumper'.
Press the
CRIS
(lower hidden switch on Model 571) and
VOL TO BE INF
switches together and wait for a click.
Press the
J
unit switch to select
MACRO
or the
K
unit switch to select
MICRO
.
014
Maximum Occlusion
Pressure
MAXIMUM
Display and/or set the maximum occlusion pressure limit. Pressure values are
displayed in
MMHG
or
CMH20
depending on the pressure units selected. See
'Selecting Pressure Units' for further information.
This pressure limit setting determines the upper limit of the pumps' variable
pressure range thereby fixing the upper limit when the
LIMIT PRES
switch is
used during normal operation.
To change the setting:
Press and hold the
PRI
(upper hidden switch on Model 571) while using the
J K
switches to adjust the value.
Press the
RUN/HOLD
switch to save the new value.
Enter Self-Test mode. See 'Entering Self-Test Mode' in previous section for instructions.
Note: For default settings refer to Appendix C 'Configured Options'.
Summary of Contents for IVAC 571
Page 1: ...IVAC Variable Pressure Volumetric Pump Models 571 572 Technical Service Manual...
Page 37: ...Chapter 4 Troubleshooting In this chapter Introduction 38 Error Messages 38 General Faults 42...
Page 97: ...Appendix C Configured Options...
Page 99: ...Appendix D Disposal...
Page 101: ...Appendix E Service Contacts...
Page 103: ...Appendix F Document History...