IVAC® Volumetric Pump (Models 571 & 572)
66/104
1000SM00018 Issue 2
Spare Parts Replacement Procedures
Body Assembly
(continued)
Separating the Front Panel Assembly from the Case
Replacement Procedure
1. Remove the two kep nuts securing the ground strap between the cover (inside top of case) and the front panel.
2. Unfasten the six retaining clips securing the front panel to the case. Separate the front panel assembly from the case.
3. Reassemble in reverse order.
Refitting note:
1) Use silicone (clear RTV coating, inner bond 140) around the case edge when refitting the front panel to the case.
Spare Parts
Item
Description
Part Number
A
FSTNR,SPD NUT LKG 0.250 SSTUD
302902
B
NUT,KEP 6-32 S PL
300036
Front Panel
(
A
) Fastener Clip (x6)
(
B
) Kep Nut (x2)
Ground Strap
To provide a clear view of the fitting area, the
illustration below shows the door removed. The door is
not
serviceable.
If a replacement door is required, contact your local
Cardinal Health Affiliate Office or Distributor for advice.