IVAC® Volumetric Pump (Models 571 & 572)
68/104
1000SM00018 Issue 2
Spare Parts Replacement Procedures
Door Assembly
Important note: The Door and the Front Panel are
not
serviceable.
If replacement of these components is required, contact your local Cardinal Health Affiliate Office or
Distributor for advice.
Bezel/Switch Panel, Display Board
Replacement Procedure
1. Remove the four screws securing the bezel/switch panel and Display Board to the door assembly.
2. Ease the bezel/switch panel and Display Board assembly away from the door until it is possible to disconnect the
interconnect flex cable connector (
H
) from the back of the Display Board (J3).
3. Remove the four caps from the Bezel ground posts.
4. Remove the three screws securing the Display Board to the bezel/switch panel.
5. Carefully separate the Display Board and the bezel/switch panel then disconnect the flex cable connector (J2) from the
Display Board.
6. Disconnect the set sensor cable harness connector (J1) and the air-in-line harness connector (J4) from the Display
Board then remove the Display Board.
7. Remove the interconnect flex cable by sliding the cable out through grommets in the door and front panel.
8. Remove the mylar shield (
L
) from the Display Board as required.
9. Reassemble in reverse order.
Introduction
To replace the components within the door assembly, follow the instructions in the relevant sections of this chapter.
These sections are:
Separating the Chassis and the Body
Bezel/Switch Panel, Display Board
Pressure Plate
Door Latch
Door Latch Pin
Air-In-Line Transmitter Assembly
Set Sensor, Pressure Disc Plunger
Pincher Cam Block
Summary of Contents for IVAC 571
Page 1: ...IVAC Variable Pressure Volumetric Pump Models 571 572 Technical Service Manual...
Page 37: ...Chapter 4 Troubleshooting In this chapter Introduction 38 Error Messages 38 General Faults 42...
Page 97: ...Appendix C Configured Options...
Page 99: ...Appendix D Disposal...
Page 101: ...Appendix E Service Contacts...
Page 103: ...Appendix F Document History...