IVAC® Volumetric Pump (Models 571 & 572)
60/104
1000SM00018 Issue 2
Spare Parts Replacement Procedures
Body Assembly
(continued)
Pincher Mechanism
Replacement Procedure
1. Remove the one screw securing the pincher mechanism to the front panel.
2. Remove the cover plate, spring and the pincher.
3. Reassemble in reverse order.
Refitting note:
Use Loctite 243 on the securing screw.
Spare Parts
Item
Description
Part Number
A
SCREW,CAP 6-32X1/4 FLH HXSO
301164
B
ADHESIVE LOCTITE 243
0000ME00672
C
COVER,PINCHER,56X/57X(BOM)
125321
D
SPRING,CAM,56X
125768
E
PINCHER,56X/57X
133248
(
A/B
) Screw / Loctite
(
C
) Cover
(
D
) Spring
(
E
) Pincher
To provide a clear view of the fitting area, the
illustration below shows the door removed. The door is
not
serviceable.
If a replacement door is required, contact your local
Cardinal Health Affiliate Office or Distributor for advice.