IVAC® Volumetric Pump (Models 571 & 572)
84/104
1000SM00018 Issue 2
Specifications
Electrical
Infusion
Rate range
1 to 999 ml/h
STANDARD RATE range
0.1 to 999.9 ml/h
ALL RATE range, with user selectable maximum rate
Volume infused range
0.1 to 9999.9 ml/h
KVO rate
5.0 ml/h or at current set rate whichever is less
Critical volume
Maximum incremental volume delivered in case of failure will not exceed 8.75 ml
System Accuracy
±5%
Infusion Set
IVAC "52" Series Volume Infusion sets
Occlusion pressure
Maximum: 500 mmHg (680 cmH20)
Mode of operation
Continuous
Battery type
Lead acid, sealed, rechargeable. Automatically charges when the pump is connected to AC
power
Battery life
5 hours @ 125 ml/h with a fully charged, new battery
Battery recharge time
Approximately 24 hours from discharge to 100% charge
Power requirements
220 to 240V AC; 50/60 Hz; 0.25A; 3-wire grounded system
Memory retention
At 25°C, with a fully charged, new battery, the volatile memory configuration information will be
retained for at least 6 months
Ground current leakage
Maximum 100µA rms (ungrounded) at 220 Vrms to meet EN/IEC 60601-1
Protection against
electrical shock hazards
Class I
Physical
Environmental
Weight
Approximately 6.5 kg
Case material
Blue enameled aluminium
Dimensions
W
H
D
12.7 cm
29.2 cm
19.3 cm
IPX rating
IPX1
Temperature
Relative humidity
Atmostpheric pressure
Operating limits
+5°C to +40°C 20% to 90% non-condensing
700 - 1060 hPa
Transport/storage limits
-40°C to +60°C 10% to 100%
500 - 1060 hPa
The following information is for reference purposes only. For more detailed specifications refer to relevant
DFU
.
Summary of Contents for IVAC 571
Page 1: ...IVAC Variable Pressure Volumetric Pump Models 571 572 Technical Service Manual...
Page 37: ...Chapter 4 Troubleshooting In this chapter Introduction 38 Error Messages 38 General Faults 42...
Page 97: ...Appendix C Configured Options...
Page 99: ...Appendix D Disposal...
Page 101: ...Appendix E Service Contacts...
Page 103: ...Appendix F Document History...