background image

1

Perfusor

®

 fm

0123

Software PFAE

Instructions for Use

Summary of Contents for Perfusor fm

Page 1: ...1 Perfusor fm 0123 Software PFAE Instructions for Use...

Page 2: ...stallation in medically used rooms must comply with the appropriate regulations e g VDE 0100 VDE 0107 or IEC publications Observe national specifications and deviations The infusion lines should be pl...

Page 3: ...roducible infusion of medication in large and small quantities The device permits extremely exacting delivery rate settings and thus optimal adaptation to any given requirements The drive system ensur...

Page 4: ...nge in position To remove syringe Pull and move to the left SF Special functions If released successive call up of Dosage Calculation Bolus function Standby function Drug selection Occlusion pressure...

Page 5: ...disinfectant Memory card window Scope of functions can be extended for exemple by way of a PCA memory card please ask for information Handle for easy transport Multi function connector MFC Connecting...

Page 6: ...atient Set delivery rate Values between 0 1 and 200 ml h Check display Correction Press C enter rate again Choose additional settings as described below For automatic rate calculation Do not enter rat...

Page 7: ...d as ml or a new target volume has to be set Target Time Time Pre selection Press key beneath TIME Enter target time via keypad and confirm with TIME 50 min 5 0 2 h 30 min 2 3 0 After confirming inste...

Page 8: ...ues are displayed Standby Pause In case of extended interruptions set values are maintained Press STOP Press key beneath SF until Standby is displayed Press key beneath ON Select special functions SF...

Page 9: ...han 2 hours is displayed after 16 hours charging time Drug Display 9 selectable drug names can be stored input via service program only Select with key beneath Drug is displayed also during infusion K...

Page 10: ...numbers with OK Entry of syringe concentration per 1 ml or per syringe filling For dosage by body weight enter body weight and confirm Otherwise confirm 0 kg With OFF complete termination of all dosag...

Page 11: ...t is switched off in Special Functions If the Perfusor fm is switched off in the mean time all previous values except the body weight are maintained Switch off from Basic Menu Press key below SF repea...

Page 12: ...s discharged Then battery alarm Yellow battery indicator flashes Red alarm indicator is on Audible alarm every 4 s Operating indicator off Staff call Switch off unit Connect to mains Syringe alarm Fau...

Page 13: ...ead in ambulance car In case of power failure or if 12 V DC 230 V AC is not connected the unit automatically will be switched to integrated rechargeable battery Charge Battery Charge battery in case o...

Page 14: ...made of syringes disposables other than those stated in the order data 10 ml Omnifix Delivery rate 1 ml h 10 ml Omnifix 5 ml h flow Pressure flow flow deviation 20 ml OPS 1 ml h deviation Max bolus v...

Page 15: ...and type of syringe 20 ml OPS Delivery rate 10 ml h flow 50 ml OPS Delivery 1 ml h flow 50 ml OPS Delivery rate 25 ml h flow 20 ml OPS 10 ml h deviation 50 ml OPS 1 ml h deviation 50 ml OPS 25 ml h de...

Page 16: ...0 4 A slow blow IEC 127 External extra low voltage 12 V Staff call Max 24 V 1 A 24 VA Arbitrary connection polarity VDE 0834 Interference suppression EMC EN 55011 CISPR 11 EN 60601 1 2 IEC 601 1 2 Ti...

Page 17: ...Manufacturer Display Type Syringe size Reference article number 10 12 ml 20 ml 50 60 ml B Braun Perfusor 8728615 8728810 Proinjekt 8728917 Omnifix 4617100 4617207 4617509 Euroject 4630505 Becton B D P...

Page 18: ...st achieved by calling in Service Individual consultation enables allowance to be made for the specific conditions in a given hospital A complete service manual can be provided if required however onl...

Page 19: ...Luer lock connectors pressure resistant 75 cm 872 2870 M PVC with loose lock nut on patient end pressure resistant 150 cm 872 2994 PE PE with Luer lock connectors pressure resistant 150 cm 872 2935 S...

Page 20: ...rawing no M 637 010 014 F04d Raw material no 3891009 8 05 01 Printed on pulp bleached 100 chlorine free B Braun Melsungen AG Postfach 1120 D 34209 Melsungen Tel 05661 71 0 Fax 05661 71 20 44 HOSPITAL...