Special functions

To stop bolus:

Press key below STOP.

Bolus without volume pre-selection

Press and hold BOL until a second BOL is

displayed.

Hold both BOL-keys. Bolus is administered

as long as both keys are pressed. Per ml bolus
administered a short audible signal sounds.

Interval bolus
Automatic bolus in set time interval.  In case of
manual bolus administration the interval bolus
is skipped.

Select bolus function under SF.

Press key below VOL, TIME or RATE. Enter

values and confirm. After value below TIME
has been confirmed, interval time runs auto-
matically.  Exit function with END, set basic
rate and start. Remaining interval time
is displayed in h:min:sec. A bolus on demand
is possible at any time during infusion at the
basal rate.

Bolus and dosage calculation

Operation as in volumetric mode. Depending

on the settings in ml (volume), quantity of
active agent (e.g. mg) or in quantity of active
agent per kg weight (mg/kg) the bolus may be
administered. During bolus administration all
three values are displayed.

Standby/Pause
In case of extended interruptions set values are
maintained.

Press STOP.

Press key beneath SF until "Standby" is

displayed.

Press key beneath ON.

Select special functions SF
Depending on the version, several functions
may be deactivated. Contact service.

To set special functions, press key below SF

repeatedly until desired special function is
displayed. – Then follow Instructions for Use as
described.

Activate Bolus Function
For additional boli.

Pressing key below ON activates the

function (values are maintained after switching
on again).

To change bolus rate:

Press key below RATE.  Enter new value

and confirm RATE. – Correction: Press C and
re-enter.

Note: Keep bolus rate low to ensure required
dosing accuracy.  Example: At a rate of
1,800 ml/h, 0.5 ml of bolus volume are
administered per second.

Exit bolus function:

Press key below END.

Bolus application during infusion:
Bolus with volume pre-selection.

Press key below BOL and release.

Display: BOLUS RELEASE?

Enter bolus. Values between 0.1 and 50.0 ml

(if no entry is made within 10 sec., function is
automatically exited).

Press key below YES. Bolus is administered.

After the bolus application the infusion
continues at basic rate.

Summary of Contents for Perfusor fm

Page 1: ...1 Perfusor fm 0123 Software PFAE Instructions for Use...

Page 2: ...stallation in medically used rooms must comply with the appropriate regulations e g VDE 0100 VDE 0107 or IEC publications Observe national specifications and deviations The infusion lines should be pl...

Page 3: ...roducible infusion of medication in large and small quantities The device permits extremely exacting delivery rate settings and thus optimal adaptation to any given requirements The drive system ensur...

Page 4: ...nge in position To remove syringe Pull and move to the left SF Special functions If released successive call up of Dosage Calculation Bolus function Standby function Drug selection Occlusion pressure...

Page 5: ...disinfectant Memory card window Scope of functions can be extended for exemple by way of a PCA memory card please ask for information Handle for easy transport Multi function connector MFC Connecting...

Page 6: ...atient Set delivery rate Values between 0 1 and 200 ml h Check display Correction Press C enter rate again Choose additional settings as described below For automatic rate calculation Do not enter rat...

Page 7: ...d as ml or a new target volume has to be set Target Time Time Pre selection Press key beneath TIME Enter target time via keypad and confirm with TIME 50 min 5 0 2 h 30 min 2 3 0 After confirming inste...

Page 8: ...ues are displayed Standby Pause In case of extended interruptions set values are maintained Press STOP Press key beneath SF until Standby is displayed Press key beneath ON Select special functions SF...

Page 9: ...han 2 hours is displayed after 16 hours charging time Drug Display 9 selectable drug names can be stored input via service program only Select with key beneath Drug is displayed also during infusion K...

Page 10: ...numbers with OK Entry of syringe concentration per 1 ml or per syringe filling For dosage by body weight enter body weight and confirm Otherwise confirm 0 kg With OFF complete termination of all dosag...

Page 11: ...t is switched off in Special Functions If the Perfusor fm is switched off in the mean time all previous values except the body weight are maintained Switch off from Basic Menu Press key below SF repea...

Page 12: ...s discharged Then battery alarm Yellow battery indicator flashes Red alarm indicator is on Audible alarm every 4 s Operating indicator off Staff call Switch off unit Connect to mains Syringe alarm Fau...

Page 13: ...ead in ambulance car In case of power failure or if 12 V DC 230 V AC is not connected the unit automatically will be switched to integrated rechargeable battery Charge Battery Charge battery in case o...

Page 14: ...made of syringes disposables other than those stated in the order data 10 ml Omnifix Delivery rate 1 ml h 10 ml Omnifix 5 ml h flow Pressure flow flow deviation 20 ml OPS 1 ml h deviation Max bolus v...

Page 15: ...and type of syringe 20 ml OPS Delivery rate 10 ml h flow 50 ml OPS Delivery 1 ml h flow 50 ml OPS Delivery rate 25 ml h flow 20 ml OPS 10 ml h deviation 50 ml OPS 1 ml h deviation 50 ml OPS 25 ml h de...

Page 16: ...0 4 A slow blow IEC 127 External extra low voltage 12 V Staff call Max 24 V 1 A 24 VA Arbitrary connection polarity VDE 0834 Interference suppression EMC EN 55011 CISPR 11 EN 60601 1 2 IEC 601 1 2 Ti...

Page 17: ...Manufacturer Display Type Syringe size Reference article number 10 12 ml 20 ml 50 60 ml B Braun Perfusor 8728615 8728810 Proinjekt 8728917 Omnifix 4617100 4617207 4617509 Euroject 4630505 Becton B D P...

Page 18: ...st achieved by calling in Service Individual consultation enables allowance to be made for the specific conditions in a given hospital A complete service manual can be provided if required however onl...

Page 19: ...Luer lock connectors pressure resistant 75 cm 872 2870 M PVC with loose lock nut on patient end pressure resistant 150 cm 872 2994 PE PE with Luer lock connectors pressure resistant 150 cm 872 2935 S...

Page 20: ...rawing no M 637 010 014 F04d Raw material no 3891009 8 05 01 Printed on pulp bleached 100 chlorine free B Braun Melsungen AG Postfach 1120 D 34209 Melsungen Tel 05661 71 0 Fax 05661 71 20 44 HOSPITAL...

Reviews: