8
Special functions
To stop bolus:
Press key below STOP.
Bolus without volume pre-selection
Press and hold BOL until a second BOL is
displayed.
Hold both BOL-keys. Bolus is administered
as long as both keys are pressed. Per ml bolus
administered a short audible signal sounds.
Interval bolus
Automatic bolus in set time interval. In case of
manual bolus administration the interval bolus
is skipped.
Select bolus function under SF.
Press key below VOL, TIME or RATE. Enter
values and confirm. After value below TIME
has been confirmed, interval time runs auto-
matically. Exit function with END, set basic
rate and start. Remaining interval time
is displayed in h:min:sec. A bolus on demand
is possible at any time during infusion at the
basal rate.
Bolus and dosage calculation
Operation as in volumetric mode. Depending
on the settings in ml (volume), quantity of
active agent (e.g. mg) or in quantity of active
agent per kg weight (mg/kg) the bolus may be
administered. During bolus administration all
three values are displayed.
Standby/Pause
In case of extended interruptions set values are
maintained.
Press STOP.
Press key beneath SF until "Standby" is
displayed.
Press key beneath ON.
Select special functions SF
Depending on the version, several functions
may be deactivated. Contact service.
To set special functions, press key below SF
repeatedly until desired special function is
displayed. – Then follow Instructions for Use as
described.
Activate Bolus Function
For additional boli.
Pressing key below ON activates the
function (values are maintained after switching
on again).
To change bolus rate:
Press key below RATE. Enter new value
and confirm RATE. – Correction: Press C and
re-enter.
Note: Keep bolus rate low to ensure required
dosing accuracy. Example: At a rate of
1,800 ml/h, 0.5 ml of bolus volume are
administered per second.
Exit bolus function:
Press key below END.
Bolus application during infusion:
Bolus with volume pre-selection.
Press key below BOL and release.
Display: BOLUS RELEASE?
Enter bolus. Values between 0.1 and 50.0 ml
(if no entry is made within 10 sec., function is
automatically exited).
Press key below YES. Bolus is administered.
After the bolus application the infusion
continues at basic rate.