17
Accuracy of set delivery rate
± 2 %, measuring time > 1 h, min. 2 ml
(including disposables approved by
manufacturer)
Delivery range
0.1 - 200.0 ml/h
Delivery pre-selection
1 - 999.9 ml (in 1 ml increments)
Occlusion alarm pressure
9-stage (approx. 0.2 bar...1.0 bar ± 0.2 bar)
Settings for bolus function
Rate
1 ml/h to 1800 ml/h
Volume
0.1 ml to 50.0 ml
Interval time 1 min. to 99 h 59 min.
Alarm in case of incorrect delivery
In case of incorrect delivery < 0.01 ml due to
apparatus malfunction the pump switches off
automatically.
Syringe table
Manufacturer
Display / Type
Syringe size / Reference article number
10/12 ml
20 ml
50/60 ml
B. Braun
Perfusor
8728615
8728810
Proinjekt
8728917
Omnifix
4617100
4617207
4617509
Euroject
4630505
Becton
B-D Plpak
300912
309661
300865
Dickinson
300912
300913
300869
300134
309663
300629
Terumo
Terumo EU
SS*10L
SS*20L
BS50LG
SS*60L
TYCO (in the past
Monoject EU
1100-6121173
1100-620036
1100-650030
Kendall/Sherwood)
Monoject US
8881-512878
8881-520657
8881-560125
* (Plastipak)
For pre-selection of syringes at the Perfusor
use reference list of article numbers. If syringes
from manufacturers are used with different
article numbers (similar) constructed than
mentioned in the above table, please contact
the original manufacturer directly.