Appendix B. Instruction 15: Control Port Serial I/O
B-9
Modes of Operation
The CR10X can read all modes of operation except the Serial ID and Signature
Analysis portion of the Test Mode. Serial input is required. The CR10X will not
read a parallel input.
Units of Output
The CR10X reads all Units of Output.
Number of Measurements Averaged
Any number of ‘measurements averaged’ may be read by the CR10X. Instruction
15 execution time is prolonged when the barometer makes several measurements
to compute an average.
Baud Rate
300 or 1200 baud
RS232 Jumpers
For typical field use where power is limited, the RS232 jumpers should be
removed to enable a TTL (0-5V) level output.
B.4.1.2 CR10X / Barometer Connection
Table B-3 CR10X/Barometer Connection Details
Table B-3 provides connection information for two TTL versions and one RS232
version of data transfer. Each configuration has specific advantages:
Configuration #1
Configuration
Configuration #3