CR10X User Guide
8-14
84 locations are allocated to Final Storage Area 2. Thus, Area 2 holds 21 seconds
(four values/second x 21 seconds = 84 locations).
When 11.4kg is exceeded, the value 10 is loaded into an input location and flag 1
is set high. The input location is then used as a down counter. The flag indicates
that an event has occurred and prevents the input location from being reloaded
until 11 seconds have passed.
The down counter is decremented by 1 each time the table is executed. When it
equals 0 all the data in Final Storage Area 2 is transferred to Final Storage Area 1
and flag 1 is set low. Instruction 96 is used with output code 81, which transfers
the entire contents of Final Storage Area 2 to Final Storage Area 1 in
chronological order, oldest to newest.
The down counter is set to 10 instead of 11 because it is decremented after
checking to see if it is 0.
Input Location Labels:
1:FORCE KG
2:DOWN CNT
*
1
Table 1 Programs
01:
1
Sec. Execution Interval
01:
P6
Full Bridge
01:
1
Rep
02:
32
7.5mV 50Hz rejection Range
03:
1
IN Chan
04:
1
Excite all reps w/EXchan 1
05:
2500
mV Excitation
06:
1
Loc [:FORCE KG]
07:
15.152 Mult
08:
0
Offset
02:
P86
Do
01:
10
Set high Flag 0 (output)
03:
P80
Set Active Storage Area
01:
2
Final Storage Area 2
02:
10
Array ID or location
04:
P77
Real Time
01:
11
Hour-Minute,Seconds
05:
P70
Sample
01:
1
Reps
02:
1
Loc FORCE KG
06:
P89
If X<=>F
01:
1
X Loc FORCE KG
02:
3
>=
03:
11.4
F
04:
30
Then Do
07:
P91
If Flag/Port
01:
21
Do if flag 1 is low
02:
30
Then Do