5
4.1.1
Connecting the accessories
Combinations of accessories that are not mentioned in the present
instructions for use may only be employed if they are specifically intended
for the respective application, and if they do not compromise the perfor-
mance and safety characteristics of the products.
Also note that any equipment connected at the interfaces must demon-
strably meet the respective IEC standards (e.g. IEC 60950 for data process-
ing equipment, IEC/DIN EN 60601-1 for electromedical devices).
All configurations must comply with basic standard IEC/DIN EN 60601-1.
Any individual connecting devices with one another is responsible for such
configuration and must ensure compliance with basic standard IEC/DIN
EN 60601-1 or applicable national standards.
►
Please address your B. Braun/Aesculap partner or Aesculap Technical
Service with any inquiries in this respect; for a contact address, see
Technical Service.
4.1.2
Presetting the correct voltage
The mains voltage must correspond to the voltage indicated on the type
plate at the back of the unit.
4.1.3
Mains power
►
Plug in the power cord at mains connection
25
of camera control
unit
7
.
►
Connect the mains plug of the power cord to the building mains.
►
Plug in the equipotentialization cable at equipotentialization connec-
tor
26
of camera control unit.
►
Connect the plug of the equipotentialization cable to the building
mains equipotentialization connector.
►
To disconnect the camera from the mains pull out the power cord from
the mains connection
25
.
4.1.4
Connecting the video cable
►
Connect the video cable to the camera rear panel and to the monitor,
documentation systems etc. Depending on the monitor and documen-
tation system use the following connections:
HD Monitor (High Definition):
■
DVI-D for digital HD 1080p quality
- or -
■
HD-SDI for digital HD 1080i/720p quality
- or -
■
RGB-Sync for analog HD 1080i/720p quality
Fig. 1
Legend
1
HD Monitor
2
SD Monitor
3
Full HD camera system
4
Mains
5
Equipotentialization
SD Monitor/SD documentation system (Standard Definition):
■
S-Video for analog SD quality
- or -
■
Video for analog SD quality
Fig. 2
Legend
1
HD Monitor
2
SD Monitor
3
Full HD camera system
4
Mains
5
Equipotentialization
6
e.g. Aesculap Eddy DVD
DANGER
Risk of injury due to unapproved configuration
using additional components!
►
For all applied components, ensure that their
classification matches that of the application
component (e.g. Type BF or Type CF) of the
respective device.
WARNING
Risk of electric shock!
►
Only connect the equipment to a supply mains
with protective earth.
1
2
3
4
5
1
2
3
4
5
6