10
Aesculap
®
Full HD 3CCD Camera system PV440
6.3
Preparations at the place of use
►
Remove any visible surgical residues to the extent possible with a
damp, lint-free cloth.
►
Transport the dry product in a sealed waste container for cleaning and
disinfection within 6 hours.
6.4
Preparation before cleaning
Camera head
►
Reprocess the product immediately after use.
►
Use suitable cleaning/disinfecting agents if the product is put away in
a wet condition.
6.5
Cleaning/disinfection
6.5.1
Product-specific safety instructions for the reprocessing pro-
cedure
6.5.2
Validated cleaning and disinfection procedure
CAUTION
Damage to the product due to improper handling
when reprocessing!
►
Separate the camera head from the camera
control unit prior to reprocessing.
►
Reprocess the camera head and the camera con-
trol unit separately.
DANGER
Risk of electric shock and fire hazard!
►
Unplug the device before cleaning.
►
Do not use flammable or explosive cleaning or
disinfecting solutions.
►
Ensure that no fluids will penetrate the product.
CAUTION
Damage to, or destruction of the product caused by
mechanical cleaning/disinfection!
►
Only clean and disinfect the product manually.
CAUTION
Damage to the product due to inappropriate clean-
ing/disinfecting agents!
►
Only use cleaning/disinfecting agents approved
for surface cleaning. Follow the manufacturer’s
instructions for the respective cleaning/disin-
fecting agent.
CAUTION
Damage to the product due to inappropriate clean-
ing/disinfecting agents and/or excessive tempera-
tures!
►
Use cleaning and disinfecting agents according
to the manufacturer’s instructions which
–
Ensure that cleaning and disinfecting agents
are approved for use on anodized aluminum,
plastics (PPSU) and stainless steel.
–
do not attack softeners (e.g. in silicone).
►
Observe specifications regarding concentration,
temperature and exposure time.
►
Do not exceed the maximum allowable temper-
ature of 60 °C.
Validated procedure
Special features
Reference
Wipe disinfection for
electrical devices with-
out sterilization
■
Camera control unit
–
Chapter Wipe disinfection for electrical devices
without sterilization
Manual cleaning with
immersion disinfection
■
Camera head
■
Suitable cleaning brush
■
20 ml disposable syringe
■
When cleaning products with movable hinges, ensure that these are
in an open position and, if applicable, move the joint while cleaning.
■
Drying phase: Use a lint-free cloth or medical compressed air
(p
max
=5 bar)
Chapter Manual cleaning/disinfecting and sub-
section:
■
Chapter Manual cleaning with immersion
disinfection