6
Aesculap
®
Full HD 3CCD Camera system PV440
4.1.5
Screwing on the endo-lens
Note
Different endo-lenses are available for the camera system.
►
Follow instructions for use of the endo-lens.
►
Screw endo-lens
1
onto camera head
5
.
4.2
Settings
4.2.1
Access to main menu
►
Unplug the camera plug
6
from the camera socket
13
.
►
Switch on camera control unit
7
.
The following test chart appears on the monitor:
►
Press and hold Push button
10
for 3 seconds.
The following message appears on the camera display:
4.2.2
Changing video formats
Note
If the aspect ratio is set to 4:3, the image is cut off on the left and right mar-
gins with SD video signals.
►
Select the camera frequency 50 Hz or 60 Hz (serial number 2000 and
higher): Put 50/60 Hz switch
18
in the corresponding position, accord-
ing to your requirements.
►
Press Push button
11
in the main menu to change the video format.
The following formats can be set:
* There is always a progressive scan signal on the DVI-D output.
4.2.3
Changing the display language
►
Press Push button
14
to change the display language.
– English
– Deutsch
– Français
– Español
– Italiano
4.2.4
Exit main menu
►
Press Push button
10
or plug in camera plug
6
.
Camera setting
Video outputs
Main menu
50/60 Hz
switch
Aspect ratio 16:9
S-Video/Video
DVI-D*
HD-SDI/
RGB
HD 1080i–
SD 16:9
50 Hz
1080p 50
1080i 50
576i PAL 16:9
HD 1080i–
SD 4:3
576i PAL 4:3
HD 720p–
SD 16:9
50 Hz
720p 50
720p 50
576i PAL 16:9
HD 720p–
SD 4:3
576i PAL 4:3
HD 1080i–
SD 16:9
60 Hz
1080p 60
1080i 60
480i NTSC 16:9
HD 1080i–
SD 4:3
480i NTSC 4:3
HD 720p–
SD 16:9
60 Hz
720p 60
720p 60
480i NTSC 16:9
HD 720p–
SD 4:3