11
6.6
Wipe disinfection for electrical devices without sterilization
RT: Room
temperature
Phase I
►
Remove any visible residues with a disposable disinfectant wipe.
►
Wipe all surfaces of the optically clean product with a fresh, disposable
disinfectant wipe.
►
Observe the specified application time (1 min minimum).
6.7
Manual cleaning/disinfecting
►
Prior to manual disinfecting, allow water to drip off for a sufficient
length of time to prevent dilution of the disinfecting solution.
►
After manual cleaning/disinfection, check visible surfaces visually for
residues.
►
Repeat the cleaning/disinfection process if necessary.
6.7.1
Manual cleaning with immersion disinfection
D–W:
Drinking water
FD–W:
Fully desalinated water (demineralized, low microbiological contamination: drinking water quality at least)
RT:
Room temperature
*Recommended: BBraun Stabimed
►
Note the information on appropriate cleaning brushes and disposable
syringes, see Validated cleaning and disinfection procedure.
Phase
Step
T
[°C/°F]
t
[min]
Conc.
[%]
Water quality
Chemical
I
Wipe disinfection
RT
≥
1
-
-
Meliseptol HBV wipes 50 % Propan-1-ol
Phase
Step
T
[°C/°F]
t
[min]
Conc.
[%]
Water
quality
Chemical
I
Disinfecting cleaning
RT (cold)
>15
2
D–W
Aldehyde-free, phenol-free, and QUAT-free concentrate, pH ~ 9*
II
Intermediate rinse
RT (cold)
1
-
D–W
-
III
Disinfection
RT (cold)
15
2
D–W
Aldehyde-free, phenol-free, and QUAT-free concentrate, pH ~ 9*
IV
Final rinse
RT (cold)
1
-
FD-W
-
V
Drying
RT
-
-
-
-