14
Aesculap
®
Full HD 3CCD Camera system PV440
8.1
Fuse replacement
Specified fuses: T 0.5 A L/250 V~; Art. no. TA020394
►
Use a small screwdriver to release the clip on the fuse holder.
►
Pull out the fuse holder.
►
Replace both fuses.
►
Reinsert the fuse holder in such a way that it audibly snaps into place.
Fig. 3
Fuse replacement
Note
If the fuses burn out frequently, the device is faulty and should be
repaired, see Technical Service.
9.
Technical Service
►
For service and repairs, please contact your national B. Braun/
Aesculapagency.
Modifications carried out on medical technical equipment may result in
loss of guarantee/warranty rights and forfeiture of applicable licenses.
Service addresses
Repairs handling camera systems:
Aesculap FLEXIMED GmbH
Repair Service
Robert-Bosch-Str. 1
79211 Denzlingen/Germany
Phone:
+49 7666 93 21 0
Fax:
+49 7666 93 21 580
E-mail:
Telephone hotline for technical assistance with electrical devices and
device combinations:
Phone:
+49 7461 95 1601
Other service addresses can be obtained from the address indicated above.
10. Accessories/Spare parts
Color bar displayed on the monitor
Camera cable not connected to the camera con-
trol unit
Connect camera cable at the camera control
unit
Inaccurate color representation on the monitor
White balance failed
Initiate white-adjustment
Inappropriate monitor color settings
Reset to factory setting
Video cable defective
Replace video cable
Malfunction
Cause
Remedy
DANGER
Risk of fatal injury from electric shock!
►
Unplug the device before changing the fuses!
WARNING
Risk of injury and/or malfunction!
►
Do not modify the product.
Art. no.
Designation
PV442
Camera head and camera cable
PV441
Camera cable
PV126S
Endo-lens with zoom
PV127S
Fixed focal length endo-lens
TA020394
Fuse set T 0.5 A L/250 V~