Error code
Cause
Remedy
E 1.1
Error in handling the probe during
measurement (contact time greater than 3
seconds)
Repeat the measurement.
Ensure correct probe handling.
Place the probe with the nececcary pressure
carefully onto the material surface.
E 1.2
Measurement cannot be evaluated (e.g. if
used probe is not suitable for test material).
Repeat measurement.
If necessary, use another probe.
E 1.3
The probe was coupled before termination
of a defined measurement time.
Repeat measurement.
Note the symbol in the MIC 10 display.
E 2.0
Error when reading the Memory Card.
Remove the Memory Card and reinsert it again.
If the error reoccurs, use another Memory Card.
E 2.1
Memory Card written by the DynaMIC
Delete data or use another Memory Card.
E 3.0
Interface is not ready.
Check to see if your printer or PC is correctly
connected.
The MIC 10 gives additional text error messages:
OFL / UFL
Overflow / Underflow
Out of range when hardness scale is
changed.
Select another hardness scale.
Full
Not enough storage capacity for the data
to be stored.
Delete the contents of the Data Logger or
Memory Card (if necessary, after previous
transfer to a PC) or use another Memory Card.
Function check and troubleshooting
Troubleshooting
9-4
Issue 08, 08/01
Krautkrämer MIC 10
Error code
Cause
Remedy
E 1.1
Error in handling the probe during
measurement (contact time greater than 3
seconds)
Repeat the measurement.
Ensure correct probe handling.
Place the probe with the nececcary pressure
carefully onto the material surface.
E 1.2
Measurement cannot be evaluated (e.g. if
used probe is not suitable for test material).
Repeat measurement.
If necessary, use another probe.
E 1.3
The probe was coupled before termination
of a defined measurement time.
Repeat measurement.
Note the symbol in the MIC 10 display.
E 2.0
Error when reading the Memory Card.
Remove the Memory Card and reinsert it again.
If the error reoccurs, use another Memory Card.
E 2.1
Memory Card written by the DynaMIC
Delete data or use another Memory Card.
E 3.0
Interface is not ready.
Check to see if your printer or PC is correctly
connected.
The MIC 10 gives additional text error messages:
OFL / UFL
Overflow / Underflow
Out of range when hardness scale is
changed.
Select another hardness scale.
Full
Not enough storage capacity for the data
to be stored.
Delete the contents of the Data Logger or
Memory Card (if necessary, after previous
transfer to a PC) or use another Memory Card.
Function check and troubleshooting
Troubleshooting
9-4
Issue 08, 08/01
Krautkrämer MIC 10
Summary of Contents for Krautkramer MIC 10
Page 6: ...0 6 Issue 08 08 01 Krautkrämer MIC 10 ...
Page 7: ...1 Introduction Krautkrämer MIC 10 Issue 08 08 01 1 1 1 Introduction ...
Page 16: ...1 10 Issue 08 08 01 Krautkrämer MIC 10 ...
Page 29: ...4 Basics of operation Krautkrämer MIC 10 Issue 08 08 01 4 1 4 Basics of operation ...
Page 37: ...5 Operation Krautkrämer MIC 10 Issue 08 08 01 5 1 5 Operation ...
Page 69: ......
Page 70: ......
Page 71: ...6 Configuration Krautkrämer MIC 10 Issue 08 08 01 6 1 6 Configuration ...
Page 88: ...7 10 Issue 08 08 01 Krautkrämer MIC 10 ...
Page 89: ...8 Care and maintenance Krautkrämer MIC 10 Issue 08 08 01 8 1 8 Care and maintenance ...
Page 97: ...10 Specifications Krautkrämer MIC 10 Issue 08 08 01 10 1 10 Specifications ...
Page 100: ...10 4 Issue 08 08 01 Krautkrämer MIC 10 ...
Page 123: ...12 Appendix Krautkrämer MIC 10 Issue 08 08 01 12 1 12 Appendix ...
Page 133: ...13 Changes Krautkrämer MIC 10 Issue 08 08 01 13 1 13 Changes ...
Page 135: ...14 Index Krautkrämer MIC 10 Issue 08 08 01 14 1 14 Index ...