You must now match the displayed value to the hard-
ness value of your test material.
–
With
^8
you can increase or decrease the dis-
played value (refer to Chapter 4, page 4-7,
Accel-
erated Setting
).
When the correct value is set, the instrument is cali-
brated for the material to be measured.
The calibration value is automatically evaluated ac-
cording to the set hardness value:
–
Press
M
.
The calibration value is displayed:
–
Note this calibration value.
If you have to carry out a new calibration for this ma-
terial you are able to enter the value directly without
having to carry out new measurements.
0
Note:
With the help of the Memory Card, recalibration with
the DL version of the MIC 10 is made easy. All calibra-
tion and adjustment parameters can be stored on a
Memory Card and loaded at any later time by inserting
this card (refer to page 5-29).
The calibration which has been changed is indicated
by the status symbol
CAL
in the measurement mode:
The calibration is retained when the instrument is
turned off.
After you have adjusted the MIC 10 to the material to
be measured you can carry out the measurement as
described above.
Operation
Hardness measurement
5-16
Issue 08, 08/01
Krautkrämer MIC 10
You must now match the displayed value to the hard-
ness value of your test material.
–
With
^8
you can increase or decrease the dis-
played value (refer to Chapter 4, page 4-7,
Accel-
erated Setting
).
When the correct value is set, the instrument is cali-
brated for the material to be measured.
The calibration value is automatically evaluated ac-
cording to the set hardness value:
–
Press
M
.
The calibration value is displayed:
–
Note this calibration value.
If you have to carry out a new calibration for this ma-
terial you are able to enter the value directly without
having to carry out new measurements.
0
Note:
With the help of the Memory Card, recalibration with
the DL version of the MIC 10 is made easy. All calibra-
tion and adjustment parameters can be stored on a
Memory Card and loaded at any later time by inserting
this card (refer to page 5-29).
The calibration which has been changed is indicated
by the status symbol
CAL
in the measurement mode:
The calibration is retained when the instrument is
turned off.
After you have adjusted the MIC 10 to the material to
be measured you can carry out the measurement as
described above.
Operation
Hardness measurement
5-16
Issue 08, 08/01
Krautkrämer MIC 10
Summary of Contents for Krautkramer MIC 10
Page 6: ...0 6 Issue 08 08 01 Krautkrämer MIC 10 ...
Page 7: ...1 Introduction Krautkrämer MIC 10 Issue 08 08 01 1 1 1 Introduction ...
Page 16: ...1 10 Issue 08 08 01 Krautkrämer MIC 10 ...
Page 29: ...4 Basics of operation Krautkrämer MIC 10 Issue 08 08 01 4 1 4 Basics of operation ...
Page 37: ...5 Operation Krautkrämer MIC 10 Issue 08 08 01 5 1 5 Operation ...
Page 69: ......
Page 70: ......
Page 71: ...6 Configuration Krautkrämer MIC 10 Issue 08 08 01 6 1 6 Configuration ...
Page 88: ...7 10 Issue 08 08 01 Krautkrämer MIC 10 ...
Page 89: ...8 Care and maintenance Krautkrämer MIC 10 Issue 08 08 01 8 1 8 Care and maintenance ...
Page 97: ...10 Specifications Krautkrämer MIC 10 Issue 08 08 01 10 1 10 Specifications ...
Page 100: ...10 4 Issue 08 08 01 Krautkrämer MIC 10 ...
Page 123: ...12 Appendix Krautkrämer MIC 10 Issue 08 08 01 12 1 12 Appendix ...
Page 133: ...13 Changes Krautkrämer MIC 10 Issue 08 08 01 13 1 13 Changes ...
Page 135: ...14 Index Krautkrämer MIC 10 Issue 08 08 01 14 1 14 Index ...