9.1 Check function
Visual check of the Vickers diamond
Periodically check the Vickers diamond under a micro-
scope.
Make sure that the pyramid-shaped diamond is not
damaged.
Checking the measurement accuracy
All UCI probes are calibrated using hardness refer-
ence plates which are certified by the material testing
institute "Materialprüfungsanstalt MPA NRW" for the
corresponding test load (1 kgf, 5 kgf, 10 kgf). Each
hardness reference plate shows natural variations that
have an influence on the measured value deviations
of the corresponding probe from the plate value.
Under well-defined conditions, as is the case with
measurements using the test support MIC 222 on
Krautkrämer hardness plates with acoustic coupling
(thin oil film or ultrasonic couplant ZG between the ref-
erence plate and base), a deviation by
±
3,6 % of the
average from 5 measurements from the plate value is
still permissible (for a max. range of 5 % with refer-
ence to the average).
Individual deviations from the plate value may occur in
freehand measurements depending on the probe
handling. However, they should not go beyond 5 %
from the plate value for 10 measurements.
You should carefully study the handling of the probe
and train yourself in working with the probe by carrying
out measurements on the hardness reference plates
until you obtain stable readings.
Repeat these reference measurements from time to
time and note down the measured average in order to
check the measurement accuracy of your probe.
Any sudden variations indicate a defective diamond or
maladjustment of the probe.
Function check and troubleshooting
Check function
9-2
Issue 08, 08/01
Krautkrämer MIC 10
9.1 Check function
Visual check of the Vickers diamond
Periodically check the Vickers diamond under a micro-
scope.
Make sure that the pyramid-shaped diamond is not
damaged.
Checking the measurement accuracy
All UCI probes are calibrated using hardness refer-
ence plates which are certified by the material testing
institute "Materialprüfungsanstalt MPA NRW" for the
corresponding test load (1 kgf, 5 kgf, 10 kgf). Each
hardness reference plate shows natural variations that
have an influence on the measured value deviations
of the corresponding probe from the plate value.
Under well-defined conditions, as is the case with
measurements using the test support MIC 222 on
Krautkrämer hardness plates with acoustic coupling
(thin oil film or ultrasonic couplant ZG between the ref-
erence plate and base), a deviation by
±
3,6 % of the
average from 5 measurements from the plate value is
still permissible (for a max. range of 5 % with refer-
ence to the average).
Individual deviations from the plate value may occur in
freehand measurements depending on the probe
handling. However, they should not go beyond 5 %
from the plate value for 10 measurements.
You should carefully study the handling of the probe
and train yourself in working with the probe by carrying
out measurements on the hardness reference plates
until you obtain stable readings.
Repeat these reference measurements from time to
time and note down the measured average in order to
check the measurement accuracy of your probe.
Any sudden variations indicate a defective diamond or
maladjustment of the probe.
Function check and troubleshooting
Check function
9-2
Issue 08, 08/01
Krautkrämer MIC 10
Summary of Contents for Krautkramer MIC 10
Page 6: ...0 6 Issue 08 08 01 Krautkrämer MIC 10 ...
Page 7: ...1 Introduction Krautkrämer MIC 10 Issue 08 08 01 1 1 1 Introduction ...
Page 16: ...1 10 Issue 08 08 01 Krautkrämer MIC 10 ...
Page 29: ...4 Basics of operation Krautkrämer MIC 10 Issue 08 08 01 4 1 4 Basics of operation ...
Page 37: ...5 Operation Krautkrämer MIC 10 Issue 08 08 01 5 1 5 Operation ...
Page 69: ......
Page 70: ......
Page 71: ...6 Configuration Krautkrämer MIC 10 Issue 08 08 01 6 1 6 Configuration ...
Page 88: ...7 10 Issue 08 08 01 Krautkrämer MIC 10 ...
Page 89: ...8 Care and maintenance Krautkrämer MIC 10 Issue 08 08 01 8 1 8 Care and maintenance ...
Page 97: ...10 Specifications Krautkrämer MIC 10 Issue 08 08 01 10 1 10 Specifications ...
Page 100: ...10 4 Issue 08 08 01 Krautkrämer MIC 10 ...
Page 123: ...12 Appendix Krautkrämer MIC 10 Issue 08 08 01 12 1 12 Appendix ...
Page 133: ...13 Changes Krautkrämer MIC 10 Issue 08 08 01 13 1 13 Changes ...
Page 135: ...14 Index Krautkrämer MIC 10 Issue 08 08 01 14 1 14 Index ...